EN
9.1 Setting the sampling time at the PX 83
On delivery the pyrometer is parameterized on a discontinuous pouring stream. If the pouring
times are < 5 seconds the measuring time needs to be adapted.
The time delay (time after detection of the pouring stream up to the start of the measurement)
and the measurement time must be less than the minimum pouring time.
Two-colour channel [
Qu
] -> ATD function [
Q.Atd
]
Parameter
Function
Default
[
T.DEL
]
Time delay [s]
1 s
(
T.ACT
)
Sampling time [s]
Auto
► Press [Mode] to access the main menu
> The display shows [
Ao
]
► Press the [▼] until [
Qu
] is displayed
► Press [Mode]
> The display shows [
Q.EPS
]
► Press [▼] until [
Q.ATD
] is displayed
► Press [Mode]
> The display shows [
t.dEl
]
► Press [▼] until [
t.Act
] is displayed
► Press [Mode]
> The current sampling time or [
Auto
] is displayed
► Press [▲ or ▼] for 2 second
> The display flashes three times
► Press [▲ or ▼] to set the desired sampling time
► Press [Mode] to confirm the parameter
Exit operation
► Wait 30 seconds
or
► Press [▲ or ▼] to change to the parameter [
End
]. Then Press [Mode] to change to the
functional menu.
► In the functional menu press [▲ or ▼] to change to the parameter [
End
], then press
[Mode].
If the time delay [
T.DEL
] must also be adjusted, proceed as described above. In this case the
parameter [
T.DEL
] must be selected and changed accordingly.
9.2 Setting the sampling time at the PX 80/81
The pyrometer is parameterized for continuous measurement on liquid metals.
The measuring time is set to 15 s in the delivery state. After the measuring time has expired, the
measured temperature is shown on the display or the analogue output. Afterwards, the measur-
ing time restarts automatically after the dead time has elapsed.
Quick Start Guide CellaCast PX 8x
Содержание ITS CellaCast PX 8 Series
Страница 1: ...CellaCast PX 8x 1109371 12 2020...
Страница 2: ......
Страница 53: ...RU CellaCast PX 8x 1 X 8x CellaCast PX www keller de its 2 3 2 1 0 630 680 4 CellaCast PX 8x...
Страница 54: ...RU 1 5 30 30 50 6 65 C 1 3 2 4 CellaCast PX 8x...
Страница 56: ...RU 8 PX xx AF xx C PX xx AF xx L 2 IO Link 2 IO Link DIN 916 DIN 911 1 CellaCast PX 8x...
Страница 60: ...RU 1 2 C 101 104 2 103 0 IO Link CellaCast PX 8x...
Страница 62: ...RU 13 KELLER HCW IO Link CellaCast PX 8x...
Страница 63: ...CN CellaCast PX 8x 1 PX 8x www keller de its 2 3 1 0mW 630 670nm 4 CellaCast PX 8x...
Страница 64: ...CN 1 5 30 30 50 6 65 C 1 3 2 4 CellaCast PX 8x...
Страница 66: ...CN PX xx AF xx C PX xx AF xx L 2 IO Link 2 IO Link DIN 911 DIN 916 1 60 C LED 2 15...
Страница 69: ...CN Mode Ao Qu Mode Q EPS Q ATD Mode t dEl t Act Mode 2 3 Mode 30 End End IO Link 10...
Страница 70: ...CN 1 2 C 101 104 103 0 IO Link 11 IO IO Link IO Link IO Link IO Link...
Страница 72: ...CN KELLER HCW IO Link CellaCast PX 8x...
Страница 74: ...KR 1 5 30 30 6 bar 50 l min 65 C 1 Target size 3 field of view 2 4 CellaCast PX 8x...
Страница 80: ...KR CellaCast PX 8x 1 Ratio correction 2 C 101 104 2 103 0 IO Link 11 IO IO IO IO IO...
Страница 82: ...KR KELLER HCW IO Link CellaCast PX 8x...
Страница 83: ......