Quick Start Guide CellaCast PX 8x
EN
To protect the laser against overload, an over temperature is provided. At temperatures above
60 °C, the laser switches of an can no longer be activated. To check, whether the laser is acti-
vated, the parameter LED lights up.
In normal operation, the laser is switch off. After activation, the laser switch of again after 2 – 15
minutes. The operator must be familiar with the pyrometer and above safety guidelines.
The activated laser pilot light can influence the measured temperature. This influence will
vary, depending on the instrument model and the temperature
If focusing is not possible during installation, focusing can also be done outside the plant.
To do this, set the pyrometer to the same distance as the distance between the pyrometer
and the pouring stream.
9 Parameterization
On delivery, the pyrometer is parameterized on a discontinuous pouring stream.
It is only necessary to change the parameters if the casting time is < 5 s
The start of a measuring cycle is automatically detected and depends on limit 1, limit 2 and the
dead time. Before starting the measurement, Limit 1 must have fallen below at least once. Limit
2 must be exceeded continuously for the duration of the dead time (T.DEL). If necessary, the
limits are to be adapted.
Two-colour channel [
Qu
] -> ATD function [
Q.Atd
]
Parameter
Function
Default
[
LI.1
]
Limit 1
1100 °C
(
LI.2
)
Limit 2
1200 °C
1: Measuring object in front of the pyrometer 4: Time delay [
T.DEL
]
2: Limit 2 [
LI. 2
]
5: Sampling time [
T.ACT
]
3: Limit 1 [
LI. 1
]
Содержание ITS CellaCast PX 8 Series
Страница 1: ...CellaCast PX 8x 1109371 12 2020...
Страница 2: ......
Страница 53: ...RU CellaCast PX 8x 1 X 8x CellaCast PX www keller de its 2 3 2 1 0 630 680 4 CellaCast PX 8x...
Страница 54: ...RU 1 5 30 30 50 6 65 C 1 3 2 4 CellaCast PX 8x...
Страница 56: ...RU 8 PX xx AF xx C PX xx AF xx L 2 IO Link 2 IO Link DIN 916 DIN 911 1 CellaCast PX 8x...
Страница 60: ...RU 1 2 C 101 104 2 103 0 IO Link CellaCast PX 8x...
Страница 62: ...RU 13 KELLER HCW IO Link CellaCast PX 8x...
Страница 63: ...CN CellaCast PX 8x 1 PX 8x www keller de its 2 3 1 0mW 630 670nm 4 CellaCast PX 8x...
Страница 64: ...CN 1 5 30 30 50 6 65 C 1 3 2 4 CellaCast PX 8x...
Страница 66: ...CN PX xx AF xx C PX xx AF xx L 2 IO Link 2 IO Link DIN 911 DIN 916 1 60 C LED 2 15...
Страница 69: ...CN Mode Ao Qu Mode Q EPS Q ATD Mode t dEl t Act Mode 2 3 Mode 30 End End IO Link 10...
Страница 70: ...CN 1 2 C 101 104 103 0 IO Link 11 IO IO Link IO Link IO Link IO Link...
Страница 72: ...CN KELLER HCW IO Link CellaCast PX 8x...
Страница 74: ...KR 1 5 30 30 6 bar 50 l min 65 C 1 Target size 3 field of view 2 4 CellaCast PX 8x...
Страница 80: ...KR CellaCast PX 8x 1 Ratio correction 2 C 101 104 2 103 0 IO Link 11 IO IO IO IO IO...
Страница 82: ...KR KELLER HCW IO Link CellaCast PX 8x...
Страница 83: ......