When installing equipment where access to the main power cord is restricted, such as 
rack mounting, a separate main input power disconnect device must be provided in 

close proximity to the equipment and within easy reach of the operator. 
For maximum safety, do not touch the product, test cables, or any other instruments 
while power is applied to the circuit under test. ALWAYS remove power from the 

entire test system and discharge any capacitors before: connecting or disconnecting 

cables or jumpers, installing or removing switching cards, or making internal changes, 

such as installing or removing jumpers. 
Do not touch any object that could provide a current path to the common side of the 

circuit under test or power line (earth) ground. Always make measurements with dry 

hands while standing on a dry, insulated surface capable of withstanding the voltage 

being measured. 
The instrument and accessories must be used in accordance with its specifications and 
operating instructions, or the safety of the equipment may be impaired. 
Do not exceed the maximum signal levels of the instruments and accessories, as defined 
in the specifications and operating information, and as shown on the instrument or test 

fixture panels, or switching card. 
When fuses are used in a product, replace with the same type and rating for continued 

protection against fire hazard. 
Chassis connections must only be used as shield connections for measuring circuits, 
NOT as protective earth (safety ground) connections. 
If you are using a test fixture, keep the lid closed while power is applied to the device 
under test. Safe operation requires the use of a lid interlock. 

 

If a screw is present, connect it to protective earth (safety ground) using the wire 

recommended in the user documentation. 

 

This symbol on an instrument means caution, risk of danger. The user should refer 
to the operating instructions located in the user documentation in all cases where 

the symbol is marked on the instrument. 

 

This symbol on an instrument means caution, risk of electric shock. Use standard 

safety precautions to avoid personal contact with these voltages. 

 

This symbol on an instrument shows that the surface may be hot. Avoid personal 
contact to prevent burns. 

 

This symbol indicates a connection terminal to the equipment frame. 

 

If the mercury symbol is on a product, it indicates that mercury is present in the 
display lamp. Please note that the lamp must be properly disposed of according to 

federal, state, and local laws. 

WARNING

  This heading in the user documentation explains dangers that might result 

in personal injury or death. Always read the associated information very 

carefully before performing the indicated procedure. 

CAUTION 

This heading in the user documentation explains hazards that could 

damage the instrument. Such damage may invalidate the warranty. 

Instrumentation and accessories shall not be connected to humans. 
Before performing any maintenance, disconnect the line cord and all test cables. 

Содержание System SourceMeter 2601B

Страница 1: ...Models 2601B 2602B and 2604B System SourceMeter Instruments Quick Start Guide ...

Страница 2: ...ment Category I and Measurement Category II as described in the International Electrotechnical Commission IEC Standard IEC 60664 Most measurement control and data I O signals are Measurement Category I and must not be directly connected to mains voltage or to voltage sources with high transient overvoltages Measurement Category II connections require protection for high transient overvoltages ofte...

Страница 3: ...y be used as shield connections for measuring circuits NOT as protective earth safety ground connections If you are using a test fixture keep the lid closed while power is applied to the device under test Safe operation requires the use of a lid interlock If a screw is present connect it to protective earth safety ground using the wire recommended in the user documentation This symbol on an instru...

Страница 4: ...or installation into a computer should never require cleaning if handled according to instructions If the board becomes contaminated and operation is affected the board should be returned to the factory for proper cleaning and servicing Power and environmental specifications For indoor use only Supply voltage range 100 V to 240 V AC 50 Hz to 60 Hz auto sensing 240 VA maximum Altitude Maximum 2000 ...

Страница 5: ...ance troubleshooting and programming commands The documentation is in PDF format on the CD ROM that is included with the instrument If you do not have Adobe Reader to view the files you can download a free copy of it at http get adobe com reader CD ROM contents The CD ROMs that are included with your instrument contain The Test Script Builder TSB Software and Series 2600B TSB add in which is a sof...

Страница 6: ...Unpack and inspect the instrument To unpack and inspect the instrument 1 Inspect the box for damage 2 Open the top of the box ...

Страница 7: ...e documentation accessories and CD ROMs 4 Carefully lift the instrument out of the box 5 Remove the packaging insert 6 Inspect the instrument for any obvious signs of physical damage Report any damage to the shipping agent immediately ...

Страница 8: ...for the Models 2601B and 2602B 4 Power line cord 5 One 8 pin screw terminal connector for the Model 2601B Two 8 Pin screw terminal connectors for the Models 2602B and 2604B 6 Series 2600B Product Information and Test Script Builder CD ROMs not shown 7 UL safety supplement not shown 8 Safety Precautions not shown 9 Models 2601B 2602B and 2604B System SourceMeter Instruments Quick Start Guide this d...

Страница 9: ...ted for dry locations Read and comply with the specifications of all instruments in the system The overall allowed signal levels may be constrained by the lowest rated instrument in the system For example if you are using a 500 V DC power supply with a 300 V DC rated switch the maximum allowed voltage in the system is 300 V DC Make sure any test fixture connected to the system protects the operato...

Страница 10: ...s 2600B can be used on a bench or in a rack Please see the instructions that came with your rack mount kit if you are installing the Series 2600B in a rack Note that the air intakes for the fan are located on the top cover and side panels of the Series 2600B The space around these areas should be free from obstruction to ensure proper fan operation Connect line power The Series 2600B operates from...

Страница 11: ...ine power 1 Make sure the front panel power switch is in the off 0 position 2 Connect the socket of the supplied power cord to the power connection on the rear panel 3 Connect the plug of the power cord to a grounded AC outlet Turn on the instrument Turn on the instrument by pressing the front panel POWER switch to the on position ...

Страница 12: ...Model 2602B the display will show KEITHLEY Model 2602B Line frequency detection and other startup checks The entire power up process takes approximately 30 seconds to complete When initialization is complete if the instrument is a Model 2602B or a Model 2604B you will see the default display screen shown below If the instrument is a Model 2601B you will see the default display screen shown below ...

Страница 13: ...test DUT for this test Step 1 Set source function range and level If the instrument has two channels Models 2602B or 2604B and is in dual channel display mode press the DISPLAY button once to show only the settings for SMU channel A 1 Press the SRC key You will see a blinking character in the SrcA value field Confirm that mV is displayed if not press the SRC key again 2 While that character is sti...

Страница 14: ...he navigation wheel to set the source value to 20 0000 V and then press the navigation wheel to enter the selection and exit EDIT mode Step 2 Set the source limit 1 Press the LIMIT key You will see a blinking character in the LimA value field 2 While that character is still blinking press the down RANGE keys as needed to select the 10 mA limit range Verify that the source limit value in the LimA f...

Страница 15: ... the Series 2600B automatically selects the best range for the measured value You will briefly see the display shown below and then the main display screen reappears Step 4 Turn the output on Turn the output on by pressing the OUTPUT ON OFF control for Model 2602B and 2604B or Model 2601B The ON OFF indicator LED lights and measurements begin Step 5 Observe measurements Observe the measured voltag...

Страница 16: ... output indicator LED turns off These steps confirm basic functionality of your Model 2601B Model 2602B or Model 2604B instrument Turn the instrument power OFF now For a basic understanding of the of the Series 2600B functionality we strongly recommend that first time users become familiar with the Series 2600B Reference Manual ...

Страница 17: ...cept Model 2400 SCPI commands For more information about loading and running the 2400 personality script see Model 2400 emulation in the Series 2600B Reference Manual What do I do if I lost or formatted the included USB flash drive How can I get another copy of the Model 2400 personality script If you lose or format the USB drive or delete the 2400 personality script you can download the latest ve...

Страница 18: ......

Отзывы: