
Test the instrument
The following test verifies basic operation of the Model 2601B,
Model 2602B, and Model 2604B. In this test, you will use the
front-panel controls shown below to source voltage and measure the
voltage output.
You do not need to connect a device under test (DUT) for this test.
Step 1: Set source function, range, and level
If the instrument has two channels (Models 2602B or 2604B) and is in
dual-channel display mode, press the
DISPLAY
button once to show
only the settings for SMU channel A.
1. Press the
SRC
key. You will see a blinking character in the
SrcA value field. Confirm that mV is displayed; if not, press the
SRC
key again.
2. While that character is still blinking, press the up or down
RANGE
keys until
40 V
is displayed.
Содержание System SourceMeter 2601B
Страница 1: ...Models 2601B 2602B and 2604B System SourceMeter Instruments Quick Start Guide ...
Страница 18: ......