Section 6: Range, Digits, Speed, Rel, and Filters
Series 2600 System SourceMeter® Instruments Reference Manual
Overview
The documentation in this section provides detailed information on characteristics and script
programming for each of the following functions:
•
•
•
•
•
Range
The selected measurement range affects the accuracy of the measurements as well as the
maximum signal that can be measured. Note that dashed lines are displayed (i.e., --.---- µA), to
indicate that the previous measurement is not recent. This usually happens when a change occurs
such as selecting a different range.
Available ranges
lists the available source and measurement ranges for the Keithley Instruments Series
2600 System SourceMeters
®
.
Table 6-1
Source and measurement ranges
Model 2601/2602
Model 2611/2612
Model 2635/2636
Voltage Ranges Current Ranges
Voltage Ranges Current Ranges Voltage Ranges
Current Ranges
100mV
1V
6V
40V
100nA
1
μ
A
10
μΑ
100
μΑ
1mA
10mA
100mA
1A
3A
200mV
2V
20V
200V
100nA
1
μ
A
10
μΑ
100
μΑ
1mA
10mA
100mA
1A
1.5A
10A
1
200mV
2V
20V
200V
100pA
2
1nA
10nA
100nA
1
μ
A
10
μΑ
100
μΑ
1mA
10mA
100mA
1A
1.5A
1. 10A range available only in pulse mode.
2. 100pA range only in measure.