2-12 TSP Programming Fundamentals
Series 2600 System SourceMeters Reference Manual
2600S-901-01 Rev. A / May 2006
Return to
The GPIB cable connectors are stackable. For additional non-Series 2600 GPIB
instruments in the test system, daisy-chain a GPIB cable from one instrument to
another.
GPIB address
At the factory, the GPIB is set for address value 26. The address value can be set
to any address value between 0 and 30. However, the address cannot conflict
with the address assigned to other instruments in the system.
The GPIB address can be changed from the communications menu. To access
the menu, press the MENU key, select COMMUNICATIONS, and then select
GPIB.
The GPIB address is saved in non-volatile memory. The address value will not
change when power is cycled or a reset command (
reset
or *RST) is sent.
RS-232
RS-232 connections
To connect the Series 2600 to the RS-232 interface, use a straight-through
RS-232 cable (see
shows the location of the RS-232
connector on the SourceMeter. When connecting the cable, make sure to tighten
the captive screws.
Figure 2-4
Male DB-9 Connector
(connect to SourceMeter)
Pin View
Female DB-9 Connector
(connect to PC)
Pin View
Straight-through RS-232 Cable
RS-232 cable (straight-through)
RS-232 settings
At the factory, the RS-232 is configured as follows:
• Baud rate: 9600
• Data bits: 8
• Parity: None
• Flow Control: None
Содержание SourceMeter 2600 Series
Страница 507: ...AppendixA Specifications Appendix A topics Model 2600S Specifications page A 2 2600S 901 01 Rev A May 2006...
Страница 575: ...AppendixF DisplayCharacterCodes Appendix F topics Introduction page F 2...
Страница 592: ......