16-4 Calibration
Series 2600 System SourceMeters Reference Manual
2600S-901-01 Rev. A / May 2006
Return to
Recommended calibration equipment
lists the recommended equipment for the calibration procedures. You
can use alternate equipment as long as that equipment has specifications equal to
or greater than those listed in the table. When possible, test equipment
specifications should be at least four times better than corresponding Series 2600
specifications.
Table 16-1
Recommended calibration equipment
Description
Manufacturer/
Model
Accuracy
Digital Multimeter
Agilent 3458A
DC Voltage
1
(2601/2602)
DC Voltage
1
(2611/2612)
DC current
1
90mV:
0.9V:
5.4V:
36V:
190mV:
1.8V:
18V:
180V:
90nA:
0.9µA:
9µA:
90µA:
0.9mA:
9mA:
90mA:
0.9A:
±8ppm
±5ppm
±4ppm
±6ppm
±5ppm
±4ppm
±6ppm
±6ppm
±430ppm
±45ppm
±25ppm
±23ppm
±20ppm
±20ppm
±35ppm
±110ppm
0.5
Ω
, 250W, 0.1%
Precision Resistor
Isotek
RUG-Z-R500-0.1-TK3
Resistance
2
0.5
Ω
:
±125ppm
50
Ω
Resistors (2)
Any suitable.
3
1. 90-day specifications show full-range accuracy of recommended model used for specified
calibration point.
2. Resistor used to calibrate Model 2601/2602 3A and 10A ranges and Model 2611/2612 1.5A and
10A ranges should be characterized to uncertainty shown using resistance function of a digital
multimeter before use.
3. Used for contact check calibration. Characterize resistors using ohms function of digital
multimeter before use.