Model 2470 High Voltage SourceMeter Instrument User's Manual
Section 5: Leakage current and insulation resistance
2470-900-01 Rev. A /
May
2019
5-5
Set up the leakage current application using the front panel
To set up the application from the front panel:
1.
Use the test leads to make the connection from the capacitor to the rear panel of the 2470, as
described in
(on page 5-2).
2.
Reset the 2470.
a.
Press the
MENU
key.
b.
Under System, select
Info/Manage
.
c.
Select
System Reset
, and then select
OK
.
3.
Press the
TERMINALS FRONT/REAR
switch to set the instrument to use the rear-panel
terminals (R is displayed to the left of the switch).
4.
Press the
HOME
key.
5.
Press the
FUNCTION
key.
6.
Under Source Voltage and Measure, select
Current
.
7.
Select the button next to Source (at the bottom of the screen).
8.
Enter
300 V
and select
OK
.
9.
Select the button next to Limit (at the bottom of the screen).
10. Enter
10 mA
and select
OK
.
11. Press the
MENU
key.
12. Under Trigger, select
Templates
.
13. Set Templates to
DurationLoop
.
14. Set Duration to a soak time of
30s
.
15. Set the delay to
0.2s
. After each reading, a delay of 0.2 s occurs; each reading is not 0.2 s apart.
16. Press the
HOME
key to return to the Home screen.
17. Press the
TRIGGER
key to turn on the output and make measurements. The output is turned off
when the measurements are complete.