Section 10: Capturing and analyzing waveforms
Model DMM7510 7½ Digit Multimeter User's Manual
10-6
DMM7510-900-01 Rev. B / May 2015
Using SCPI commands
You may need to make changes so that this code will run in your programming environment.
Send the following commands:
Command
Description
*RST
:SENS:DIG:FUNC "VOLT"
:SENS:DIG:VOLT:RANG 10
:SENS:DIG:VOLT:INP AUTO
:SENS:DIG:VOLT:SRATE 1000000
:SENS:DIG:VOLT:APER AUTO
:SENS:DIG:COUN 100
:TRAC:POIN 100
:TRAC:CLE
:TRAC:TRIG:DIG
:TRAC:DATA? 1, 100, "defbuffer1", READ
:TRAC:DATA? 1, 100, "defbuffer1", REL
•
Reset the Model DMM7510.
•
Set the instrument to measure digitize voltage.
•
Set range to 10 V.
•
Set the input impedance to 10 G
Ω
.
•
Set the sample rate to 1e6.
•
Enable automatic aperture.
•
Set the sample count to 100.
•
Set
defbuffer1
to store 100 points.
•
Clear
defbuffer1
.
•
Digitize the signal.
•
Read the first 100 measurement and relative
timestamp values from
defbuffer1
.
Using TSP commands
The following TSP code is designed to be run from Keithley Instruments Test Script Builder (TSB).
TSB is a software tool that is available from the Keithley Instruments website. You can install and
use TSB to write code and develop scripts for TSP-enabled instruments. Information about how to
use TSB is in the online help for TSB and in the “Introduction to TSP operation” section of the
Model
DMM7510 Reference Manual
.
To use other programming environments, you may need to make changes to the example TSP code.
By default, the Model DMM7510 is configured to use the SCPI command set. You must select the
TSP command set before sending TSP commands to the instrument.
To enable TSP commands:
1. Press the
MENU
key.
2. Under System, select
Settings
.
3. For Command Set, select
TSP
.
4. At the prompt to reboot, select
Yes
.