Section 3: Using a remote interface
Model DMM7510 7½ Digit Multimeter User's Manual
3-14
DMM7510-900-01 Rev. B / May 2015
Identify the instrument
If you have a bank of instruments, you can click the ID button to determine which one you are
communicating with.
To identify the instrument:
In the middle of the left side of the Home page, click the
ID
button.
The button turns green and the LAN status indicator on the instrument blinks.
Click the
ID
button again to return the button to its original color and return the LAN status indicator to
steady on.
Review events in the event log
The event log records all LXI events that the instrument generates and receives. The log includes the
following information:
•
The EventID column, which shows the event identifier that generated the event.
•
The System Timestamp column, which displays the seconds and nanoseconds when the event
occurred.
•
The Data column, which displays the text of the event message.
To clear the event log and update the information on the screen, click the
Refresh
button.
Determining the command set you will use
You can change the command set that you use with the Model DMM7510. The remote command sets
that are available include:
•
SCPI: An instrument-specific language built on the SCPI standard.
•
TSP: A scripting programming language that contains instrument-specific control commands that
can be executed from a stand-alone instrument. You can use TSP to send individual commands
or use it to combine commands into scripts.
You cannot combine the command sets.
As delivered from Keithley Instruments, the Model DMM7510 is set to work with the Model
DMM7510 SCPI command set.
To set the command set from the front panel:
1.
Press the
MENU
key.
2.
Under System, select
Settings
.
3.
Select the button next to Command Set.
4.
Select the command set.
5.
You are prompted to reboot.