Model 8011 Connection Instructions
8
PA-1032 Rev. A / April 2011
Table 1. Connections for two-terminal device test (local sense)
Cable end
Cable
Cable end
Model 2651A rear panel
CA-552-1 (supplied with the Model 2651A)
Open
LO of CA-552-1
Black CA-560-0
8010-120 blue terminal 4
HI of CA-552-1
Red CA-560-2
8010-120 yellow terminal 6
Figure 5. Connections for two-terminal device test (local sense)
HI
CA-560-0
(black)
2-T
erminal DUT
Plug
HIGH CURRENT TEST SOCKET KIT
8011
RATINGS:
40V MAX.
15A DC MAX.
50A PULSE MAX.
From
Model 2651A
CA-552-1
(supplied with
Model 2651A)
8010-317 insulator plug
(insert into 3-pin socket)
CA-560-2
(red)
TO-247 Device Test
Board (8010-120)
Two-terminal DUT using a Model 2651A set to remote sense
Tools needed
•
Small flat-head screwdriver
Instructions
1. Remove power from the Model 2651A.
2. Make connections as shown in Table 2 and Figure 6.
3. Install the 8010-317 insulator plug into the test socket.
4. Install the 2-terminal DUT into the axial posts labeled 4, 1, 3 and 6 on the 8010-120 board.
5. Return power to the Model 2651A.