manualshive.com logo in svg
background image

 

Model 8011 Connection Instructions 

 

Specifications are subject to change without notice. 

PA-1032 Rev. A / April 2011 

15 

 

Maintenance personnel

 perform routine procedures on the product to keep it operating properly, for example, setting the line 

voltage or replacing consumable materials. Maintenance procedures are described in the user documentation. The procedures 
explicitly state if the operator may perform them. Otherwise, they should be performed only by service personnel. 

Service personnel

 are trained to work on live circuits, perform safe installations, and repair products. Only properly trained 

service personnel may perform installation and service procedures. 

 

Keithley Instruments products are designed for use with electrical signals that are rated Measurement Category I and 
Measurement Category II, as described in the International Electrotechnical Commission (IEC) Standard IEC 60664. Most 
measurement, control, and data I/O signals are Measurement Category I and must not be directly connected to mains voltage or 
to voltage sources with high transient over-voltages. Measurement Category II connections require protection for high transient 
over-voltages often associated with local AC mains connections. Assume all measurement, control, and data I/O connections 
are for connection to Category I sources unless otherwise marked or described in the user documentation. 

Exercise extreme caution when a shock hazard is present. Lethal voltage may be present on cable connector jacks or test 
fixtures. The American National Standards Institute (ANSI) states that a shock hazard exists when voltage levels greater than 
30 V RMS, 42.4 V peak, or 60 VDC are present. A good safety practice is to expect that hazardous voltage is present in any 
unknown circuit before measuring. 

Operators of this product must be protected from electric shock at all times. The responsible body must ensure that operators 
are prevented access and/or insulated from every connection point. In some cases, connections must be exposed to potential 
human contact. Product operators in these circumstances must be trained to protect themselves from the risk of electric shock. 
If the circuit is capable of operating at or above 1000V, no conductive part of the circuit may be exposed. 

 

Do not connect switching cards directly to unlimited power circuits. They are intended to be used with impedance-limited 
sources. NEVER connect switching cards directly to AC mains. When connecting sources to switching cards, install protective 
devices to limit fault current and voltage to the card. 

Before operating an instrument, ensure that the line cord is connected to a properly-grounded power receptacle. Inspect the 
connecting cables, test leads, and jumpers for possible wear, cracks, or breaks before each use. 

When installing equipment where access to the main power cord is restricted, such as rack mounting, a separate main input 
power disconnect device must be provided in close proximity to the equipment and within easy reach of the operator. 

For maximum safety, do not touch the product, test cables, or any other instruments while power is applied to the circuit under 
test. ALWAYS remove power from the entire test system and discharge any capacitors before: connecting or disconnecting 
cables or jumpers, installing or removing switching cards, or making internal changes, such as installing or removing jumpers. 

 

Do not touch any object that could provide a current path to the common side of the circuit under test or power line (earth) 
ground. Always make measurements with dry hands while standing on a dry, insulated surface capable of withstanding the 
voltage being measured. 

The instrument and accessories must be used in accordance with its specifications and operating instructions, or the safety of 
the equipment may be impaired. 

Do not exceed the maximum signal levels of the instruments and accessories, as defined in the specifications and operating 
information, and as shown on the instrument or test fixture panels, or switching card. 

When fuses are used in a product, replace with the same type and rating for continued protection against fire hazard. 

Chassis connections must only be used as shield connections for measuring circuits, NOT as safety earth ground connections. 

If you are using a test fixture, keep the lid closed while power is applied to the device under test. Safe operation requires the use 
of a lid interlock. 

If a 

 screw is present, connect it to safety earth ground using the wire recommended in the user documentation. 

The 

!

 symbol on an instrument means caution, risk of danger. The user should refer to the operating instructions located in 

the user documentation in all cases where the symbol is marked on the instrument. 

The 

 symbol on an instrument means caution, risk of electric shock. Use standard safety precautions to avoid personal 

contact with these voltages. 

The 

  symbol on an instrument shows that the surface may be hot. Avoid personal contact to prevent burns. 

The   symbol indicates a connection terminal to the equipment frame. 

Содержание 8011

Страница 1: ... SourceMeter instruments and the Model 2651A High Power SourceMeter Instrument at currents up at 50 A pulsed and voltages up to 40 V The Model 8011 includes two test boards the 8011 DTB and the 8011 CTB and cables and connectors for simplifying connections between the SourceMeter instruments and the device The 8011 DTB Device Test Board is designed for testing type TO 247 packaged parts and axial ...

Страница 2: ...er use Keithley Model number 8010 DTB Customizable Test Board 8010 130 to re order use Keithley Model number 8010 CTB PA 1032 Model 8011 High Current Test Socket Kit Connection Instructions this document Four nylon hex standoffs with inserts on each end ST 272 1 Four Phillips flat head screws 6 32 x 3 8 Two 8 in 203 2 mm stackable banana cables one black CA 560 0 one red CA 560 2 One mounting plat...

Страница 3: ...ropriate for your application 2 Connect one end of the ground cable to the ground screw at the center of the mounting plate 3 Attach the ground cable with the 6 32 Keps washer nut 4 Thread the four nylon hex standoffs onto the four threaded posts on the mounting plate 5 Place the 8010 120 or 8010 130 board on the standoffs 6 Thread four of the Phillips flat head screws through the 8010 120 or 8010...

Страница 4: ... 6 32 x 3 8 Phillips flat head screws 8010 130 Customizable Test Board or 8010 120 TO 247 Device Test Board Four ST 272 1 hex nylon standoffs One 8011 301 mounting plate One CA 568 120 green yellow ground cable with lugs One 6 32 Keps washer nut Connect ground cable lug to protective earth ground ...

Страница 5: ...nting plate should be used to secure your test board 8010 120 or 8010 130 It can also be used to secure the assembled Model 8011 to a work surface using the mounting holes shown in Figure 3 Figure 3 8011 301 mounting plate dimensions 156 in 3 962 mm 5 250 in 133 35 mm 2 625 in 66 675 mm 750 in 19 05 mm 4 500 in 114 3 mm 6 000 in 152 4 mm Mounting holes ...

Страница 6: ...ng fixture to the 8010 130 Customizable Test Board using the 6 32 screw mounts provided Figure 4 8010 130 Customizable Test Board dimensions 3 000 in 76 2 mm 1 000 in 25 4 mm 156 in 3 962 mm 6 32 750 in 19 05 mm 3 500 in 88 9 mm 5 000 in 5 mm 250 in 6 35 mm 4 250 in 107 95 mm 3 750 in 95 25 mm 1 750 in 44 45 mm ...

Страница 7: ...ons Use cases are shown below that demonstrate some typical connections Use cases Two terminal DUT using a Model 2651A set to local sense NOTE These instructions provide general information about connections to the Model 2651A For detail on the connections see the Model 2651A user documentation Tools needed Small flat head screwdriver Instructions 1 Remove power from the Model 2651A 2 Make connect...

Страница 8: ...l DUT Plug HIGH CURRENT TEST SOCKET KIT 8011 RATINGS 40V MAX 15A DC MAX 50A PULSE MAX From Model 2651A CA 552 1 supplied with Model 2651A 8010 317 insulator plug insert into 3 pin socket CA 560 2 red TO 247 Device Test Board 8010 120 Two terminal DUT using a Model 2651A set to remote sense Tools needed Small flat head screwdriver Instructions 1 Remove power from the Model 2651A 2 Make connections ...

Страница 9: ... 2651A Open LO of CA 552 1 Black CA 560 0 8010 120 blue terminal 4 HI of CA 552 1 Red CA 560 2 8010 120 yellow terminal 6 Model 2651A rear panel CA 557 1 supplied with the Model 2651A CS 1629 8 Sense LO of CS 1629 8 Black CA 567 0 8010 120 blue terminal 1 Sense HI of CS 1629 8 Red CA 567 2 8010 120 yellow terminal 3 Figure 6 Connections for two terminal device test remote sense S LO G G G G S HI 2...

Страница 10: ...el System SourceMeter Instrument Model 2612A Dual Channel System SourceMeter Instrument 200 V Tools and equipment needed Small flat head screwdriver Cable created using the 2600 KIT screw terminal connector kits included with the Model 8011 the cable must use wire that can support 3 A and 200 V Instructions 1 Remove power from the SourceMeter instruments 2 Make connections as shown in Table 3 and ...

Страница 11: ...567 2 8010 120 white terminal 2 Sense LO of CS 1629 8 Black CA 567 0 8010 120 yellow terminal 3 LO of CS 1629 8 Black CA 567 0 8010 120 yellow terminal 3 Sense HI of CS 1629 8 Red CA 567 2 8010 120 blue terminal 1 HI of CS 1629 8 Red CA 567 2 8010 120 blue terminal 4 Figure 7 Connections for three terminal device test G D S LO S LO G HI G G G S HI S LO G G G G S HI HI 2600 KIT 3 Terminal DUT HIGH ...

Страница 12: ...s NOTE This use case calls for three 2600 ALG 2 cables Normally the Model 2635A ships with one of these cables and the Model 2636A ships with two For additional cables order Keithley Instruments catalog number 2600 ALG 2 This cable is listed as part number 237 308 2 on Model 2635A and 2636A parts lists To set up a test of a three terminal device 1 Remove power from the SourceMeter instruments 2 Ma...

Страница 13: ... terminal 6 HI of CA 552 1 Red CA 560 2 8010 120 white terminal 5 Sense LO of CS 1629 8 Black CA 567 0 8010 120 yellow terminal 3 Sense HI of CS 1629 8 Red CA 567 2 8010 120 white terminal 2 Figure 8 Connections for three terminal device test L O b l a c k Sense LO red G D S S LO G G G G S HI HI Model 2651A CA 567 2 red CA 560 0 black CA 560 2 red CS 1629 8 CA 552 1 supplied with Model 2651A CA 55...

Страница 14: ... explosion from causing harm The following safety precautions should be observed before using this product and any associated instrumentation Although some instruments and accessories would normally be used with nonhazardous voltages there are situations where hazardous conditions may be present This product is intended for use by qualified personnel who recognize shock hazards and are familiar wi...

Страница 15: ...rds directly to AC mains When connecting sources to switching cards install protective devices to limit fault current and voltage to the card Before operating an instrument ensure that the line cord is connected to a properly grounded power receptacle Inspect the connecting cables test leads and jumpers for possible wear cracks or breaks before each use When installing equipment where access to th...

Страница 16: ...acks must be purchased from Keithley Instruments Standard fuses with applicable national safety approvals may be used if the rating and type are the same Other components that are not safety related may be purchased from other suppliers as long as they are equivalent to the original component note that selected parts should be purchased only through Keithley Instruments to maintain accuracy and fu...

Отзывы: