Model 8011 Connection Instructions
10
PA-1032 Rev. A / April 2011
Three-terminal DUT using a Model 2651A and a single or dual channel
SourceMeter instrument
NOTE
If you are using a Model 2600A, do not set the output above 40 V. The Model 8011 is rated to 40 V.
You can use the Model 8011 with the 8010-120 board to test a three-terminal device using a Model 2651A High
Power System SourceMeter
®
Instrument and another SourceMeter instrument. This use-case is appropriate for the
following SourceMeter instruments:
•
Model 2601A Single-Channel System SourceMeter Instrument
•
Model 2611A Single-Channel System SourceMeter Instrument (200 V)
•
Model 2602A Dual-Channel System SourceMeter Instrument
•
Model 2612A Dual-Channel System SourceMeter Instrument (200 V)
Tools and equipment needed
•
Small flat-head screwdriver
•
Cable created using the 2600-KIT screw terminal connector kits (included with the Model 8011); the cable must
use wire that can support 3 A and 200 V
Instructions
1. Remove power from the SourceMeter instruments.
2. Make connections as shown in Table 3 and Figure 7.
3. Install the three-terminal DUT into the socket on the 8010-120 board.
4. Return power to the SourceMeter instruments.
5. On both SourceMeter Instruments:
a. Press
the
CONFIG
key.
b. Press
the
SRC
or
MEAS
key (the sense mode can be set from either the V-SOURCE or the V-MEAS
menu).
c. If you pressed the SRC key, select
V-SOURCE > SENSE-MODE
.
d. If you pressed the MEAS key, select
V-MEAS > SENSE-MODE
.
e. Select
4-WIRE
to enable remote sense.
6. Test the device.