Multiplexer card connections
Test equipment connection diagrams in this section are generic in nature and are intended
only to show basic connections on how to connect test equipment to the relay contact(s) under
test. See Section 2 for complete details on card connections and terminal designations.
Channel resistance tests
Perform the following steps to verify that each contact of every relay is closing properly and
that the channel resistance is within specification.
1.
Turn on the Model 2010 DMM, and allow it to warm up for one hour before making
measurements.
2.
Set the Model 2010 to the 10
Ω
range, and connect the four test leads to the INPUT and
SENSE
Ω
4 WIRE jacks.
3.
Short the free ends of the four test leads together, and enable REL on the Model 2010 to
null out residual resistance. Leave REL enabled for the entire test.
4.
Connect the Model 2010 INPUT and SENSE
Ω
4 WIRE jacks to the relay contact to be
tested, as shown in Figure 4-1. Note that 4-wire connections to the card terminals are
used.
5.
Close the relay being tested.
6.
Note the resistance reading on the Model 2010, and verify that it is <1
Ω
.
7.
Open the relay being tested.
8.
Repeat steps 4 through 7 for every relay contact on the card.
2010 MULTIMETER
RANGE
!
F
500V
PEAK
FRONT/REAR
3A 250V
AMPS
HI
LO
INPUTS
350V
PEAK
1000V
PEAK
SHIFT
LOCAL
POWER
RANGE
R
SHIFT
CH1
REM
TALK
LSTN
SRQ
STAT
REL
FILT
4W
BUFFER
MATH
REAR
SCAN
TIMER
STEP
CH2
CH3
CH4
CH5
CH6
CH7
CH8
CH9
CH10
HOLD
TRIG
FAST
MED
SLOW
AUTO
ERR
INPUT
SENSE
Ω
4 WIRE
AUTO
EXIT
ENTER
DIGITS RATE
REL
FILTER
TRIG
EX TRIG
STORE RECALL
OPEN CLOSE
DCV
DCI
MX+B
%
dBm
ACV
ACI
Ω
2
Ω
4
FREQ
TEMP
dB
CONT
PERIOD SENSOR
LIMITS
ON/OFF
DELAY
HOLD
SAVE
SETUP
CONFIG
HALT
TYPE
RS232
GPIB
RATIO
STEP
SCAN
DRYCKT O COMP
CAL
TEST
Model 2010 DMM
Connect to INPUT and SENSE
In
Out
H
L
G
Close
Relay
Model 7999-2 Relay
Under Test. See Section 2
for Detailed Connections.
Figure 4-1
Channel resistance
test connections
Содержание 7999-2
Страница 1: ...Contains Operating and Servicing Information Model 7999 2 Multiplexer Card Instruction Manual ...
Страница 12: ...1 GeneralInformation ...
Страница 16: ...2 Connections ...
Страница 23: ...3 OperatingConsiderations ...
Страница 28: ...4 ServiceInformation ...
Страница 36: ...5 ReplaceableParts ...
Страница 39: ......
Страница 40: ......
Страница 41: ......
Страница 42: ......
Страница 44: ...Keithley Instruments Inc 28775 Aurora Road Cleveland Ohio 44139 Printed in the U S A ...