Channel-to-channel isolation tests
CAUTION
The following procedure uses the Model 6517A voltage source to measure
resistance. To avoid damage to the Model 7999-2, use the Model 6517A
manual ohms mode to limit the test voltage to 30V as outlined in the proce-
dure.
Perform the following steps to check channel-to-channel isolation:
1.
Turn on the Model 6517A Electrometer, and allow the unit to warm up for one hour
before testing.
2.
Select the Model 6517A ohms function and 2G
Ω
range, and enable zero check. Make
sure the voltage source is in standby.
3.
Connect the electrometer to the two Model 7999-2 channels being tested, as shown in
Figure 4-3. Note that electrometer INPUT HI is connected to one channel, while
VOLTAGE SOURCE HI is connected to the other channel.
4.
Put the Model 6517A in the manual ohms mode, and set the voltage source to +30V.
5.
Place the Model 6517A voltage source in operate, and disable zero check.
6.
Allow the reading to settle, then verify that the resistance reading is >1G
Ω
.
7.
Enable zero check, and place the Model 6517A voltage source in standby.
8.
Repeat steps 3 through 7 for all other two-channel combinations on the card.
WARNING:
NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
WARNING:
NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
CAUTION:
FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
CAUTION:
FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
INPUT
250V PEAK
!
LINE RATING
50-60HZ
50VA MAX
AC ONLY
LINE FUSE
SLOWBLOW
1/2A 90-125V
1/4A 180-250V
IEEE-488
(CHANGE IEEE ADDRESS
WITH FRONT PANEL MENU)
DIGITAL
I/O
TRIG LINK
1
15V
LO
HI
RS232
!
MADE
IN
U.S.A.
PREAMP OUT
250 PEAK
COMMON
2V ANALOG
OUTPUT
TEMP
TYPE K
HUMIDITY
OPTION SLOT
V SOURCE
EXT TRIG
IN
MTR COMP
OUT
INTERLOCK
SELECTED
LINE VOLTAGE
90-110V
105-125V
180-220V
210-250V
1010V
PEAK
750V
PEAK
Model 6517A Electrometer
In
Out
H
L
G
Close
Relay
Model 7999-2 Relays Under Test. See
Section 2 for Detailed Connections.
Connect Voltage
Source HI to
Second Relay
In
Out
H
L
G
Close
Relay
INPUT
Voltage Source HI
Connect INPUT HI
to First Relay
Figure 4-3
Channel-to-channel
isolation connections
Содержание 7999-2
Страница 1: ...Contains Operating and Servicing Information Model 7999 2 Multiplexer Card Instruction Manual ...
Страница 12: ...1 GeneralInformation ...
Страница 16: ...2 Connections ...
Страница 23: ...3 OperatingConsiderations ...
Страница 28: ...4 ServiceInformation ...
Страница 36: ...5 ReplaceableParts ...
Страница 39: ......
Страница 40: ......
Страница 41: ......
Страница 42: ......
Страница 44: ...Keithley Instruments Inc 28775 Aurora Road Cleveland Ohio 44139 Printed in the U S A ...