8
4 x 4 Non-blocking matrix example
Using a Model 2750 and four Model 7711/7712 cards, a 4
×
4 non-blocking matrix can be created.
Figure 5
shows how to
connect the 7711/7712 modules. This configuration provides a convenient way to simultaneously connect up to 4 DUTs to 4
different test instruments. Use a short coaxial cable to connect the OUTs of each module as shown. contains a couple of
sample commands to control the matrix.
To use , select the IN connection (IN 1 – IN 4), and then in the same square select the OUT connection (OUT1 - OUT4). Send
the
ROUTe:MULTiple:CLOSe
command with the channels in both “Ch” columns.
Table 3
4
××××
4 Sample non-blocking matrix commands
Command
Signal path
ROUTe:MULTiple:CLOSe (@101,405)
Routes signal from IN1(OUT A of slot 1) through
OUT4 (OUT B of slot 4).
ROUTe:MULTiple:CLOSe (@101,405,108,302)
Routes signal from IN1 (OUT A of slot 1) through
OUT4 (OUT B of slot 4), and also from IN2 (OUT B
of slot 1) through OUT1 (OUT A of slot 4).
Table 4
4
××××
4 Non-blocking matrix channels
IN
Ch
Ch
OUT
IN
Ch
Ch
OUT
IN 1
101
405
OUT4
IN 3
201
407
OUT4
102
401
OUT3
202
403
OUT3
103
305
OUT2
203
307
OUT2
104
301
OUT1
204
303
OUT1
IN
Ch
Ch
OUT
IN
Ch
Ch
OUT
IN 2
105
406
OUT4
IN 4
205
408
OUT4
106
402
OUT3
206
404
OUT3
107
306
OUT2
207
308
OUT2
108
302
OUT1
208
304
OUT1