4-6
Four-terminal ohms measurements
For general purpose testing, the Keithley Model 196
can be used to make four-terminal resistance measure-
ments of each thick
Þ
lm. As shown in Figure 4-5,
OHMS HI and OHMS SENSE HI are connected to one
matrix row, and OHMS LO and OHMS SENSE LO are
connected to another matrix row. With this con
Þ
gura-
tion, the resistance of each resistor element and/or
combined elements can be measured by closing the ap-
propriate crosspoints. In Figure 4-5, crosspoints 1!1
(Row 1, Column 1) and 2!3 (Row 2, Column 3) are
closed to measure the combined resistance of R1 and
R2.
Figure 4-4
Thick film resistor network testing
1
2
3
4
H
1
3
4
5
6
7
8
9
10
2
R1
R2
R3
R4
R1
R2
R4
R3
Model 230
Model 196
Ohms Sense
Volts Ohms
Volts/Ohms HI
Output
Common
Measure V or
4-terminal
Ω
Source V
Ohms Sense HI
Volts/Ohms LO
Ohms Sense LO
TF-1
TF-2
Sense Output
Sense Common
L
H L
H L
H L
H L
H L
H L
H L
H L
H L
Rows
H
L
H
L
H
L
H
L
Cols
7012
The effects of thermal EMFs generated by relay con-
tacts and connections can be cancelled by using the off-
set compensated ohms feature of the Model 196. (The
Model 7012 has been designed to keep relay EMF at a
minimal level.) To compensate for thermal EMFs, close
two crosspoints (such as 1!1!1 and 1!2!1), this will short
the input of the Model 196, enable zero to cancel inter-
nal offset, and then enable offset compensated ohms.
Artisan Scientific - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisan-scientific.com