![Keithley 7002 Скачать руководство пользователя страница 20](http://html2.mh-extra.com/html/keithley/7002/7002_quick-reference-manual_4024208020.webp)
16
The scan can be terminated by sending the following com-
mand:
:ABORt
All closed channels will open and the Model 7002 will
return to the idle state.
Channel patterns (STORE and RECALL)
Up to 500 channel patterns can be stored in memory. A
channel pattern is simply a pattern of open and closed chan-
nels.
There are two methods to store a channel pattern. One
method is to store an actual open and closed pattern of
relays (RELAY PATTERN) and the other method is to store
the channel list (CURRENT LIST).
When a RELAY PATTERN is recalled, each channel will
assume the state determined by the stored relay pattern.
When a CURRENT LIST is recalled, the channels con-
tained in the stored Channel List will close. Channels not
included in the list will open.
Storing a Relay Pattern:
1. Close the desired channels.
2. Press STORE and select RELAY PATTERN.
3. Key in the memory location (1 to 500) and press
ENTER.
Artisan Scientific - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisan-scientific.com