15

 

IEEE-488 Bus Commands —

 

 Over the bus, the :CLOSe 

and :OPEN commands are used to control channels. Exam-
ples: 

:CLOSe (@ 4!1, 4!5)

Close channels 1 and 5 of Slot 4.

:OPEN (@ 4!1, 4!5)

Open channels 1 and 5 of Slot 4. 

:OPEN ALL

Open all channels. 

 

Scanning channels

 

Channels are scanned in the order that they are presented in 
the Scan List. To perform a scan, first program the control-
ling aspects of the scan (see Table 4), and then press (or 
click) STEP to take the Model 7002 out of the idle state 
(ARM indicator on). 

Regardless of how the scan is configured, the STEP key (or 
LED) is always active allowing you to manually step 
through the scan. The scan can be terminated at any time by 
pressing (or clicking) OPEN ALL. All closed channels will 
open and the Model 7002 will return to the idle state (ARM 
indicator off). 

 

IEEE-488 bus commands —

 

 Over the bus, a scan is con-

figured using the Trigger Subsystem commands (see Table 
16). The following command starts the scan by taking the 
Model 7002 out of the idle state: 

:INITiate 

Artisan Scientific - Quality Instrumentation ... Guaranteed  |  (888) 88-SOURCE  |  www.artisan-scientific.com

Содержание 7002

Страница 1: ...cess underutilized and idle equipment along with credit for buybacks and trade ins Custom engineering so your equipment works exactly as you specify Critical and expedited services Leasing Rentals Dem...

Страница 2: ...Model 7002 Switch System Quick Reference Guide Artisan Scientific Quality Instrumentation Guaranteed 888 88 SOURCE www artisan scientific com...

Страница 3: ...k Reference Guide 1993 Keithley Instruments Inc All rights reserved Cleveland Ohio 44139 U S A Document Number 7002 903 01 Rev B 4 00 Artisan Scientific Quality Instrumentation Guaranteed 888 88 SOURC...

Страница 4: ...ation Summary 4 Menu Structures 16 Default Conditions 21 Error and Status Messages 24 Status Registers 27 IEEE 488 2 Common Commands 31 SCPI Command Subsystems 33 Artisan Scientific Quality Instrument...

Страница 5: ...se of the instrument They must be protected from electric shock and contact with hazardous live circuits Maintenance personnel perform routine procedures on the product to keep it operating for exampl...

Страница 6: ...o switching cards install protective devices to limit fault current and voltage to the card Before operating an instrument make sure the line cord is connected to a properly grounded power receptacle...

Страница 7: ...n The symbol on an instrument indicates that the user should refer to the operating instructions located in the manu al The symbol on an instrument shows that it can source or measure 1000 volts or mo...

Страница 8: ...the product If you are unsure about the applicability of a replacement component call a Keithley In struments office for information To clean an instrument use a damp cloth or mild water based cleane...

Страница 9: ...ailed information consult the Model 7002 Instruction Manual All Keithley product names are trademarks or registered trademarks of Keithley Instruments Inc Other brand and product names are trademarks...

Страница 10: ...ect TYPE SLOT 1 through SLOT 10 Use the cursor keys and to check card type assign ment for any slot To change a slot assignment select the slot use the cursor keys to change the card type and press EN...

Страница 11: ...ontrol the LED Graticule Disbling the LED Graticule eliminates the dimly lit state for LEDs LEDs are either off or brightly lit See LED Graticule Control for more details Relay status display mode In...

Страница 12: ...e on or off of each available channel in your switch system Changing the display mode The display mode can be changed through the MAIN MENU Press MENU Select GENERAL Select DISPLAY Select SHOW LIST RE...

Страница 13: ...ticule control is performed from the MAIN MENU as follows Press MENU Select GENERAL Select DISPLAY Select GRATCULE CONTROL ON OFF Select OFF to disable the LED Graticule or ON to enable it Table 1 Sig...

Страница 14: ...aults The instrument will return to the fac tory default conditions listed in Table 5 on power up if the Power On Default is set for Factory Default as follows List On Channel not available Channel no...

Страница 15: ...ve the setup in a memory loca tion 0 9 Use the RESTORE selection to return the instru ment to the configuration saved at the specified memory location The instrument will return to user saved defaults...

Страница 16: ...he Model 7002 to default conditions RST Return 7002 to RST defaults listed in Table 5 SYSTem PRESet Return 7002 to SYSTem PRESet defaults listed in Table 6 The Model 7002 can be programmed to power on...

Страница 17: ...he list Use the hyphen key to specify a channel range i e 1 1 1 10 B Light Pen Method Click the corresponding channel LED to enter the channel into the list Use the hyphen LED to specify a channel ran...

Страница 18: ...r to the end of the list IEEE 488 bus commands Over the bus the Channel List is included with the commands to open and close chan nels see Closing and Opening Channels The Scan List is created using t...

Страница 19: ...Model 7002 out of the idle state ARM indicator on Regardless of how the scan is configured the STEP key or LED is always active allowing you to manually step through the scan The scan can be terminate...

Страница 20: ...d closed pattern of relays RELAY PATTERN and the other method is to store the channel list CURRENT LIST When a RELAY PATTERN is recalled each channel will assume the state determined by the stored rel...

Страница 21: ...IEEE 488 bus commands Over the bus the following commands are used to store and recall channel patterns MEMory SAVE M num Save Relay Pattern at specified memory location num 1 to 500 MEMory SAVE LIST...

Страница 22: ...RESET Setup Menu Save setup at a memory loca tion up to 10 Return 7002 to setup stored at a memory location Power on Menu Power on to factory default setup conditions Power on to setup stored at a mem...

Страница 23: ...Display messages in English Display messages in Ger man Display messages in French GENERAL SERIAL STATUS MESSAGES DISPLAY LED DISPLAY SETUP SHOW LIST GRATICULE CON TROL BREAK BEFORE MAKE General Menu...

Страница 24: ...structure Menu item Description CHAN CONTROL CHANNEL SPACING TIMER EXTERNAL GPIB MANUAL TRIGLINK ASYNCHRONOUS SEMI SYNCHRONOUS Configure Channel Layer Select channel spacing Use a timer to select eac...

Страница 25: ...e Source Bypass Disable the Source Bypass SCAN CONTROL SCAN SPACING TIMER EXTERNAL GPIB MANUAL TRIGLINK IMMEDIATE HOLD Configure Scan Layer Select scan spacing Use the timer to control scan spacing Us...

Страница 26: ...e a bus trigger to arm scanner Use an external trigger to arm scanner Use a Trigger Link trig ger to arm scanner Use to hold up the scan in the arm layer Define number of times to arm scanner Continuo...

Страница 27: ...erlock Break Before Make LED Graticule Relay status All open Cleared No change No change High No change Off No change Off 0 sec Manual STEP key Scan list length Source Immediate Infinite Acceptor Imme...

Страница 28: ...Ne OFF 1 IMMediate ACCeptor 1 2 1 0 IMMediate 0 001 ACCeptor 1 2 OFF 1 IMMediate ACCeptor 1 2 INFinite 0 IMMediate 0 001 ACCeptor 1 2 TRIGger SEQuence1 COUNt AUTO DELay SOURce TIMer TCONfigure PROToco...

Страница 29: ...to 5 SCHannel DISPlay SMESsage WINDow3 GRATicule ON 0 OFF OFF OFF ON ON 0 OFF OFF OFF ON Table 6 RST and SYStem PRESet Defaults cont Command name RST value SYSTem PRESet value Artisan Scientific Quali...

Страница 30: ...ntification error EE Saved state error EE 350 174 Too many channels closed EE Re entering the idle layer SE 173 172 171 161 126 Waiting in arm layer 2 SE Waiting in arm layer 1 SE Waiting in trigger l...

Страница 31: ...data error EE Invalid character data EE Character data too long EE Character data not allowed EE String data error EE 151 154 158 160 161 Invalid string data EE String too long String data not allowed...

Страница 32: ...330 Cannot create program EE Illegal program name EE Program currently running EE Program syntax error EE Self Test failed EE 350 410 420 430 440 Queue overflow EE Query interrupted EE Query untermin...

Страница 33: ...On URQ User Request CME Command Error EXE Execution Error DDE Device Dependent Error QYE Query Error OPC Operation Complete Logical AND OR Logical OR OPC PON B7 URQ B6 CME B5 EXE B4 DDE B3 QYE B2 B1 B...

Страница 34: ...ister PTR Transition NTR Filter Operation Event Register Operation Event Enable Register Idle B10 Arm B6 Trig B5 Set B1 B15 B11 B9 B8 B7 B4 B3 B2 B0 Idle B10 Arm B6 Trig B5 Set B1 B15 B11 B9 B8 B7 B4...

Страница 35: ...TR Positive Transition Register NTR Negative Transition Register 0 B14 B10 B9 B15 0 Chan B14 B10 B9 B15 0 Chan PTR NTR Questionable Transition Filter Questionable Event Register B14 B10 B9 B15 To Ques...

Страница 36: ...Request Enable Register OSB Operation Summary Bit MSS Master Summary Status IEEE 488 2 RQS Request for Service IEEE 488 1 ESB Event Summary Bit MAV Message Available QSB Questionable Summary Bit EAV E...

Страница 37: ...Request the pro grammed value of the Standard Event Status Enable Regis ter Request the pro grammed value of the Standard Event Status Register and clear it Request the manu facturer model num ber se...

Страница 38: ...on Sets the contents of the Service Request Enable Register The Model 7002 returns the value of the Service Request Enable Register Returns the value of the Status Byte Reg ister This command issues a...

Страница 39: ...ckets in the program message 2 Angle brackets are used to indicate parameter type Do not use angle brackets in the program message 3 Upper case characters indicate the short form version for each comm...

Страница 40: ...ode 0 or 1 Path to locate message to bottom dis play Path to control user text messages Define ASCII message a using up to 32 characters Query text message Enable 1 or ON or disable 0 or OFF message m...

Страница 41: ...line 1 Path to set polarity of digital output line 2 Select polarity active high AHIGh or active low ALOW Query polarity of line 2 Path to set polarity of digital output line 3 Select polarity active...

Страница 42: ...n 0 closed Define scan list Query number of channels in scan list Query scan list Specify channels that can not be closed Query channels that cannot be closed Specify interlock X 1 to 5 Define List 1...

Страница 43: ...ect pole mode 1 2 or 4 for specified slot Query pole mode for specified slot Specify delay 0 to 99999 999 sec for specified slot Query delay for specified slot Path to program memory Path to save chan...

Страница 44: ...el b LEVel TTL3 LEVel b LEVel TTL4 LEVel b LEVel Digital output line 1 Set line true 1 or ON or false 0 or OFF Query state of digital output line Digital output line 2 Set line true 1 or ON or false 0...

Страница 45: ...ister Query positive transition register Program the negative transition register Query negative transition register Query condition register Path to control arm event registers Query event register P...

Страница 46: ...h to control trigger event registers Query event register Program the enable reg ister Query enable register Program the positive transition register Query positive transition register Program the neg...

Страница 47: ...cify error and status messages for queue Query enabled messages Specify messages not to be placed in queue Query disabled messages Table 15 SYSTem command summary Command Description SYSTem PRESet POS...

Страница 48: ...program arm layer 1 Immediately go to Layer2 Program arm count 1 to 9999 or INF Query arm count Select control source HOLD IMMediate MANual BUS TLINk EXTernal Query control source Loop around control...

Страница 49: ...ct input line 1 to 6 Query input line Select output line 1 to 6 Query output line Path to program trigger layer Immediately perform a scan step Program measure count 1 to 9999 or INF Enable 1 or ON or...

Страница 50: ...Enable SOURce or disable ACCeptor Bypass Query direction Path to configure asynchro nous Trigger Link Select input line 1 to 6 Query input line Select output line 1 to 6 Query output line Path to con...

Страница 51: ...Keithley Instruments Inc 28775 Aurora Road Cleveland Ohio 44139 Printed in the U S A Artisan Scientific Quality Instrumentation Guaranteed 888 88 SOURCE www artisan scientific com...

Страница 52: ...uipment Have surplus equipment taking up shelf space We ll give it a new home Learn more Visit us at artisantg com for more info on price quotes drivers technical specifications manuals and documentat...

Отзывы: