6517B

Electrometer/High Resistance Meter

IEEE-488 BUS IMPLEMENTATION

IMPLEMENTATION:

 SCPI (IEEE-488.2, SCPI-1999.0).

TRIGGER TO READING DONE:

 150ms typical, with external trigger.

RS-232 IMPLEMENTATION: Supports:

 SCPI 1991.0. 

Baud Rates:

 300, 600, 1200, 2400, 

4800, 9600, 19.2k, 38.4k, 57.6k, and 115.2k.

FLOW CONTROL:

 None, Xon/Xoff.

CONNECTOR:

 DB-9 TXD/RXD/GND. 

GENERAL

OVERRANGE INDICATION:

 Display reads “OVERFLOW” for readings >105% of range. The 

display reads “OUT OF LIMIT” for excesive overrange conditions.

RANGING:

 Automatic or manual.

CONVERSION TIME:

 Selectable 0.01PLC to 10PLC.

MAXIMUM INPUT:

 250V peak, DC to 60Hz sine wave; 10sec per minute maximum on 

mA ranges.

MAXIMUM COMMON MODE VOLTAGE (DC to 60Hz sine wave): 

Electrometer, 500V peak; 

V Source, 750V peak.

ISOLATION (Meter COMMON to chassis):

 >10

10

Ω

, <500pF.

INPUT CONNECTOR:

 Three lug triaxial on rear panel.

2V ANALOG OUTPUT:

 2V for full range input. Non-inverting in Volts mode, inverting when 

measuring Amps, Ohms, or Coulombs. Output impedance 10k

Ω

.

PREAMP OUTPUT:

 Provides a guard output for Volts measurements. Can be used as an 

inverting output or with external feedback in Amps and Coulombs modes. 

EXTERNAL TRIGGER:

 TTL compatible External Trigger and Electro meter Complete.

GUARD:

 Switchable voltage guard available.

DIGITAL I/O AND TRIGGER LINE:

 Available, see manual for usage.

EMC: 

Conforms to European Union Directive 89/336/EEC, EN 61326-1.

SAFETY: 

Conforms to European Union Directive 73/23/EEC, EN 61010-1.

READING STORAGE:

 50,000.

READING RATES:

 

To Internal Buffer:

 425 readings/second

1

.  

To IEEE-488 Bus:

 400 readings/second

1, 2

.  

Bus Transfer: 

3300 readings/second

2

.

1. 0.01PLC, digital filters off, front panel off, tempe RH off, Line Sync off.
2. Binary transfer mode.

DIGITAL FILTER:

 Median and averaging.

ENVIRONMENT:

 

Operating:

 0°–50°C; relative humidity 70% non-condensing, up to 35°C. 

Storage:

 –25° to +65°C.

ALTITUDE:

 Maximum 2000 meters above sea level per EN 61010-1.

WARM-UP:

 1 hour to rated accuracy (see manual for recommended procedure).

POWER:

 User selectable 100, 120, 220, 240VAC ±10%; 50/60Hz, 100VA max.

PHYSICAL: Case Dimensions:

 90mm high 

×

 214mm wide 

×

 369mm deep (3½ in. 

×

 8½ in. 

×

 14½ in.).

Working Dimensions:

 From front of case to rear including power cord and IEEE-488 

connector: 15.5 inches.

Net Weight: 

5.4kg (11.8 lbs.).

Shipping Weight:

 6.9kg (15.11 lbs.).

COULOMBS

ACCURACY

TEMPERATURE 

  (1 Year)

1, 2

 

COEFFICIENT 

6½-DIGIT 

18°–28°C 

0°–18°C & 28°–50°C 

RANGE RESOLUTION 

±(% + offset) 

±(% + offset)/°C

  2 nC 

1  fC 

0.4  +  50  fC 

0.04 +  30  fC

 20 nC 

  10  fC 

0.4  + 500  fC 

0.04 + 100  fC

200  nC 

100  fC 

0.4  +  5 pC 

0.04 +  1  pC

 2 µC 

1 pC 

0.4  +  50 pC 

0.04 +  10  pC

NOTES

1. Specifications apply immediately after charge acquisition. Add

 

|Q

AV

(4fA + _____  ) T

A

 

 RC
where T

A

 = period of time in seconds between the coulombs zero and meas urement and 

Q

AV

 = average charge measured over T

A

, and RC = 300,000 typical.

2. When properly zeroed, 6½-digit, 1PLC (power line cycle), median filter on, digital filter = 10 readings.

INPUT BIAS CURRENT:

 <4fA at T

CAL

. Temperature coefficient = 0.5fA/°C, 2nC range.

TEMPERATURE (Thermocouple)

ACCURACY (1 Year)

1

 

THERMOCOUPLE

18°–28°C

TYPE  

RANGE 

±(% rdg + °C)

–25°C to 150°C

±(0.3% + 1.5°C)

NOTES

1. Excluding probe errors, T

cal

 ± 5°C, 1 PLC integration time.

HUMIDITY

ACCURACY (1 Year)

1

 

RANGE 

18°–28°C, ±(% rdg + % RH)

0–100% 

±(0.3% +0.5)

NOTES

1. Humidity probe accuracy must be added. This is ±3% RH for Model 6517-RH, up to 65°C probe environment, 

not to exceed 85°C.

6517B rear panel.

65

17