Model 6517A Getting Started Manual
Remote Operation
3-13
:RESistance
Path to configure Ohms:
(Volts commands)
The commands for Volts (except :GUARd)
apply.
[:AUTO]
Path to configure Auto V-Source Ohms:
:RANGe
Path to set measurement range:
[:UPPer] <n>
Select range (0 to 100e18).
:AUTO <b>
Enable or disable auto range.
:MANual
Path to configure Manual V-Source Ohms:
:CRANge
Path to set measurement range:
[:UPPer] <n>
Select range (0 to 100e18).
:AUTO <b>
Enable or disable auto range.
:VSOurce
Path to control the V-Source.
[:AMPLitude] <n>
Specify V-Source level (0 to 1000).
:RANGe <n>
Set range;
≤
100 = 100V range, >100 =
1000V range.
:OPERate <b>
Enable or disable V-Source output.
:IREFerence <b>
Enable or disable amps reference.
:DAMPing <b>
Enable or disable damping.
:VSControl <name>
Select ohms V-Source mode (MANual or
AUTO).
:MSELect <name>
Select ohms type (NORMal or RESistivity).
:RESistivity
Path to configure resistivity:
:STHickness <NRf>
Volume — Specify sample thickness; 0.0001
to 99.9999 (mm).
:FSELect <name>
Select test fixture (M8009 or USER).
:M8009
Path to query Model 8009 test fixture:
:RSWitch?
Query switch setting (SURFace or
VOLume).
:USER
Path to configure user test fixture.
:RSELect <name>
Select test type (SURFace or VOLume).
:KSURface <NRf>
Surface — Specify Ks; 0.001 to 999.999.
:KVOLume <NRf>
Volume — Specify Kv; 0.001 to 999.999.
:CHARge
Path to configure Coulombs:
(Volts commands)
The commands for Volts (except :GUARd)
apply.
:ADIScharge
Path to control auto discharge.
:LEVel <NRf>
Set level (-2.2e6 to 2.2e6).
[:STATe] <b>
Enable or disable auto discharge.