3-17
3.8
Calibration errors
3.8.1 Error query commands
Purpose
To request calibration errors over the bus.
Format
:cal:prot:eerr?
Request cal execution error status
:cal:prot:verr?
Request voltage function cal errors
:cal:prot:aerr?
Request current function cal errors
:cal:prot:cerr?
Request charge function cal errors
:cal:prot:terr?
Request temperature function cal errors
:cal:prot:herr?
Request humidity function cal errors
:cal:prot:serr?
Request voltage source cal errors
:cal:prot:ferr?
Request factory cal errors
Parameter
None
Description
The calibration error queries allow you to access the various calibration error messages that
might be generated during calibration. A separate query is provided for each function or cate-
gory. For example, the :cal:prot:verr? query is used to request only those errors that occur while
calibrating the voltage function.
Programming Note
See Appendix B for a complete listing of calibration error messages.
Examples
:CAL:PROT:AERR?
Request amps cal errors.
:CAL:PROT:SERR?
Request voltage source cal errors.