3-14
Measurements and Sourcing Voltage
Model 6485 and 6487 User’s Manual
Open interlock indication
If the interlock is asserted (opened) while the unit is on the 50V or 500V range, the voltage
source will also technically be in compliance. However, there will be no indication of that
status over the front panel or in the status registers. The open interlock takes precedence.
SCPI commands — voltage source
Table 3-7
SCPI commands — voltage source
Command
Description
SOUR:VOLT <Voltage>
SOUR:VOLT:RANG <Range>
SOUR:VOLT:ILIM <Limit>
SOUR:VOLT:STAT <b>
SOUR:VOLT:INT <b>
SOUR:VOLT:INT:FAIL?
Set the source output level (-500 to +500V).
Set source range to one of three ranges: 10, 50, and 500V.
Set the source current limit: 2.5e-5, 2.5e-4, 2.5e-3, or 2.5e-2.
1
Enable (ON) or disable (OFF) source output.
Enable (ON) or disable (OFF) interlock for 10V range.
2
Query interlock state (1 = asserted); source output cannot be turned on.
1
2.5e-2 not available for 50V and 500V ranges.
2
See
Programming example
— voltage
The following command sequence will output 5V on the 10V range with a 2.5mA limit:
*RST
' Return 6487 to GPIB defaults.
SOUR:VOLT:RANG 10
' Select 10V source range.
SOUR:VOLT 5
' Set voltage source output to 5.
SOUR:VOLT:ILIM 2.5e-3
' Set current limit to 2.5mA.
SOUR:VOLT:STAT ON
' Put voltage source in operate.
Содержание 6485
Страница 40: ...1 28 Getting Started Model 6485 and 6487 User s Manual...
Страница 94: ...7 8 Remote Operation and Commands Model 6485 and 6487 User s Manual...
Страница 95: ...A Specifications...
Страница 98: ...A 4 Specifications Model 6485 and 6487 User s Manual...
Страница 105: ...C ExamplePrograms Programming examples lists and defines three types of programming exam ples...
Страница 113: ......
Страница 114: ......