C-2
Data Flow
Introduction
Data
fl
ow for remote operation is summarized by the block diagram shown in Figure C-1.
Refer to this block diagram for the following discussion.
The SENSE block represents the basic measured readings of voltage, current and resistance.
If Filter is enabled, the readings will be
fi
ltered. The SENSE block also measures time for the
timestamp.
When the INITiate command is sent, the programmed number of source-measure operations
are performed and the respective data is temporarily stored in the Sample Buffer. For example,
if 20 source-measure operations were performed, then 20 sets of data will be stored in the Sam-
ple Buffer. Data from this buffer is then routed to other enabled data
fl
ow blocks.
SENSE
(Measurements)
Volts, Amps, Ohms,
Timestamp, Filter
Sample
Buffer
CALC1
Math
Expression
CALC2
Limit Tests
NULL (Rel)
Trace
(Data Store)
FETCh?
READ?
MEAS? CALC1:DATA?
CALC2:DATA?
CALC3:DATA?
TRACe:DATA?
CALC3
Min, Max, Sdev
Mean, Pk-Pk
Figure C-1
Data flow block
diagram
Содержание 6430
Страница 26: ......
Страница 32: ......
Страница 78: ...2 14 Connections ...
Страница 98: ...3 20 Basic Source Measure Operation ...
Страница 138: ...5 30 Source Measure Concepts ...
Страница 156: ...6 18 Range Digits Speed and Filters ...
Страница 168: ...7 12 Relative and Math ...
Страница 176: ...8 8 Data Store ...
Страница 202: ...9 26 Sweep Operation ...
Страница 248: ...11 22 Limit Testing ...
Страница 310: ...16 6 SCPI Signal Oriented Measurement Commands ...
Страница 418: ...17 108 SCPI Command Reference ...
Страница 450: ...18 32 Performance Verification ...
Страница 477: ...A Specifications ...
Страница 489: ...B StatusandErrorMessages ...
Страница 498: ...B 10 Status and Error Messages ...
Страница 499: ...C DataFlow ...
Страница 503: ...D IEEE 488BusOverview ...
Страница 518: ...D 16 IEEE 488 Bus Overview ...
Страница 519: ...E IEEE 488andSCPI ConformanceInformation ...
Страница 523: ...F MeasurementConsiderations ...
Страница 539: ...G GPIB488 1Protocol ...
Страница 557: ......