background image

b.  To  check  these  resistors,  it  is  necessary  to  use 

a  bridge  capable  of  better  than  1%  accuracy  up  to  lOll 
ohms  such  as  the  Keithley  Instruments  Mode15l5A Megohm 
Bridge.  If  such  equipment  is  not  available,  two  pro­

cedures  are  recommended  to  check  out  the  resistors: 

1.  Return  t�e  complete  instrument  to  the  factory 

for  resistor  calibration. 

2.  Replace  the  high-megohm  resisters  periodically 

with  a  certified  set  from  Keithley  Instruments  to 

assure  absolute  calibration  accuracy. 

6-11. 

ACCURACY  CHECK. 

a. 

Checking  the  accuracy  is  the  quickest  way  to 

spot  faulty  Model  6l0C  operation. 

Perform  the  check 

about  once  a  year,  if  components  are  replaced,  or  if 

other  adjustments  are  made.  If  accuracy  is  verified 

over  all  ranges,  the  Model  6l0C  should  be  able  to  meet 
all  specifications.  If  the  accuracy  must  be  checked 
often,  check  the  stability. 

b. 

Connect  the  Model  6l0C  input  to  the 

Model  241  and  monitor  the  output  with  the  Model  160 
DVM. 

First,  set  the  Model  6l0e  for  the  10-volt  range. 

Increase  the  input  voltage  in  I-volt  steps  from  0  to 
10  volts.  The  Model  6l0e  should  indicate  the  input 
voltage  to  ±l%  of  full  scale. 

Perform  the  same  oper­

ation  in  the  3-volt  range.  Connect  the  Model  610e  to 
the  Model  260. 

Check  the  other  voltage  ranges  for 

accuracy  of  ±l%  of  full  scale. 

Make  sure  the  3V-lMA 

Switch  is  in  the  3V  position  for  the  1,  3  and  10  volt 
ranges 

NffiE 

In  subparagraph  c  below,  for  10-1  ampere 
through  10-5  ampere  ranges,  a  larger  Multi­
plier  Switch  setting  will  result  in  inaccu­
racies  due  to  loading  of  the  current  source; 
hence  the  .01  Multiplier  Switch  setting  must 

be  used.  The  appropriate  current  source  is 

two  decades  less  than  range  setting;  that  is, 
on  the  10-1  ampere  range  use  10-3  ampere 

source,  etc. 

c. 

Current.  Connect  the  Model  6l0e  to  the  Model 

261  with  the  Model  2611  and  monitor  the  output  with 
the  Model  160  DVM. 

Check  the  full-scale  accuracy  of 

all  the  current  \,ositions  on  the  Range  Switch.  For 
the  10-6  to  10-1  ampere  ranges,  set  the  FEEDBACK 
Switch  to  FAST  and  the  Multiplier  Switch  to  1.  The 

Model  610C  should  indicate  the  input  current  to  =2%  of 
full  scale  from  10-6  to  10-8  am\,ere  ranges  and  ±4%  of 
full  scale  from  the  10-9  to  10- 1  ampere  ranges. 

For 

the  ranges  above  10-5  ampere,  construct  a  current 
source  whose  output  is  ±0.2S%.  Set  the  Model  610C 
FEEDBACK  Switch  to  NORMAL  and  the  Multiplier  Switch  to 

.01  when  calibrating  these  ranges. 

26 

d.  Ohms.  Connect  the  hig�  side  of  the  Model  5155 

Resistor-to  the  center  terminal  of  the  INPUT  Receptacle 
and  the  low  side  to  low.  Check  the  full  scale  accuracy 
of  all  ohm  pOSitions  on  the  Range  Switch  by  measuring 

the  voltage  drop  of  the  known  current  across·  the  known 
sample  with  the  Model  6l0C.  The  test  current  applied 
is  the  reciprocal  of  the  OHMS  Range  Setting. 

Check 

the 

Ion 

through  108  ohm  ranges  for  an  accuracy  of  ±3% 

of  full  scale  by  using  the  resistance  measurement  meth­

od  as  described  in  paragraph  2-5b 

Switch  in 

NORMAL). 

Check  the  109  through 

ohm  ranges  for  an 

accuracy  of  ±5%  of  full  scale  by  employing  the  resis­

tance  measurement  method  as  described  in  paragraph  2-Sc 

(FEEDBACK  Switch  in  FAST). 

e. 

Coulombs. 

To  check  the  Model  610e  as  a  charge 

amplifier,  set  the  FEEDBACK  Switch  to  FAST.  Apply  10 
volts  from  the  Keithley  Model  241  Voltage  Supply 
through  a  Keithley  Model  5155  lOll  ohm  standard  to  the 
Model  6l0C  input.  Set  the  Multiplier  Switch  to  the 
settings  given  in  Table  9. 

Use  a  stop  watch  or  a  Tek­

tronix  Model  561A  Oscilloscope  to  time  the  rise  to 
full  scale  deflection.  Check  each  range  to  ±5%. 

TABLE  9. 

Coulomb  Ranges  Accuracy  Check.  The  Table  gives  the 
Multiplier  Switch  setting,  the  coulomb  range  checked 

and  the  rise  time  for  each  coulomb  range. 

Multiplier  Switch 

Setting 

.01 
.1 

10 

6-12. 

UNITY  GAIN  CHECK. 

Coulombs 

Range 

Rise  Time,  Zero 

to  Full  scale 

(seconds) 

10 
10 
10 
10 

a.  Zero  check  the  Model  610e  and  set.the  instrument 

to  the  100  volt  range. 

Connect  the  Model  241  Voltage 

Supply  to  the  Model  610C  INPUT  Receptacle  and  to  the 
Model  662  Differential  Voltmeter  High  terminal.  Con­

nect  the  Electrometer 

Xl 

OUTPUT  Terminal  to  the  Low 

terminal  of  the  Model  662. 

Set  the  Model  662  dial  to 

and  the  Null  Switch  to  10  millivolts. 

The  Model  662 

should  indicate  at  or  near  zero. 

NOTE 

A  large  transient  is  generated  when  the  241 

is  switched  from  Standby  to  Output.  To  pre­
vent  false  indication,  set.241  to  Output  and 
use  voltage  dial  (0  or  10  volts)  when  making 
this  check. 

24

Содержание 610C

Страница 1: ...Model 610C 610CRSolid State Electrometers Instruction Manual A G R E A T E R M E A S U R E O F C O N F I D E N C E...

Страница 2: ...RECT SPECIAL INCIDENTAL OR CONSEQUENTIAL DAMAGES ARISING OUT OF THE USE OF ITS INSTRUMENTS AND SOFTWARE EVEN IF KEITHLEY INSTRUMENTS INC HAS BEEN ADVISED IN ADVANCE OF THE POSSIBILITY OF SUCH DAMAGES...

Страница 3: ...ates that a shock hazard exists when voltage levels greater than 30V RMS 42 4V peak or 60VDC are present A good safety practice is to expect that hazardous voltage is present in any unknown circuit be...

Страница 4: ...components in mains circuits including the power transformer test leads and input jacks must be purchased from Keithley Instru ments Standard fuses with applicable national safety approvals may be use...

Страница 5: ...Model 610C 610CR Solid State Electrometers Instruction Manual Keithley Instruments Inc All rights reserved Cleveland Ohio U S A Third Printing December 2001 Document Number 29113 Rev C...

Страница 6: ...i CONTENTS Section Page SPECIFICATIONS ii 1 GENERAL DESCRIPTION 1 2 OPERATION 4 3 CIRCUIT DESCRIPTION 9 4 ACCESSORIES 12 5 SERVICING 15 6 CALIBRATION 24 7 REPLACEABLE PARTS 29 SCHEMATIC 37 13 22 27 35...

Страница 7: ...ric Measurements Resistance Measurements Recorder Output 100 mV Divider Network Voltmeter Configuration NORMAL Mode Ammeter Configuration FAST Mode Ohmmeter Configuration NORMAL Mode Chassis Side View...

Страница 8: ...opposite input polarity GAIN 0 1 0 3 etc to 3000 FREQUENCY RESPONSE within 3db DC to 40kHz at a gain of I and lower decreasing to DC to 100Hz at maximum gain Full output response limited to 3kHz on an...

Страница 9: ...ess than 1 millivolt per 24 hours Zero offset due to temperature change is less than 150 microvolts per C after a 30 minute warm up This offset how ever can easily be compensated for with the front pa...

Страница 10: ...used with AMPERES OHMS or COULOMBS the setting should be multiplied by the RANGE setting Controls power to instrument Also selects the meter polarity and center scale Meter zero adjustment Meter zero...

Страница 11: ...EDIUM 5101 SWITCH S107 RANGE SWITCH 5106 ZERO CHECK SWITCH S103 ZERO CONTROL POWER COARSE S102 NEON FEEDBACK SWITCH 5105 GROUND J102 INPUT J101 DS101 FIGURE 3 Front Panel Controls Model 6l0eR 1MA CAL...

Страница 12: ...he FEEDBACK Switch at FAST some insulators such as Teflon may produce ran 10m signals which show up as erratic meter deflections nsulation used in the Model 6l0C is carefully select zd to minimize the...

Страница 13: ...e re ciprocal value of the AMPERES setting For instance to obtain an input resistance of 10 megohms 107n set the RANGE switch to the 10 7 AMPERES range Then adjust the Multiplier switch to the full sc...

Страница 14: ...CABLE CONNECT UNKNOWN BETWEEN HIGH AND GND r I I I FIGURE 5 Guarded Voltage Measurement T IS ES RS ERROR IS 1M ERROR EIN ES x 100 IF EIN ES THEN 1 M IS FIGURE 6 Current Measurements Loading Error 1 V...

Страница 15: ...be The circuit connections are shown in Figure 8 mine the resistance the Multiplier setting subtracted from the voltage source setting resistance is therefore the ratio of the net divided by the curre...

Страница 16: ...r MA CAL MODEL 610C I I I I I I I I J NOTE In the FAST mode the OUTPUT is floating with respect to chassis ground 2 Current Output 1 milliampere for full scale input variable 5 with 1400 ohm recorder...

Страница 17: ...of the Model 6l0C is as follows UHF INPUT 1 The amplifier is always in a unity gain input voltage to output current converter configuration The internal circuitry is arranged such that a fu11 scale in...

Страница 18: ...f input capacity it also reduces the input drop to less than 100 microvolts 10 3 5 OIlMl ETER OPERATION a Normal Method In the normal method of resis tance measurements FEEDBACK Switch in NORMAL posi...

Страница 19: ...UHF INPUT 11 A OUTPUT RANGE MULTIPLIER GUARD FIGURE 11 Ammeter Configuration FAST Mode xl OUTPUT UHF INPUT OUTPUT MULTIPLIER FIGURE 12 Ohmmeter Configuration NORMAL Mode II...

Страница 20: ...ading Model 6105 Resistivity Chamber The Model 6105 is a guarded test fixture for measurement of surface and volume resistivity The chamber is designed in accordance with ASTM Standard Method of Test...

Страница 21: ...ace components only as necessary Use only reli able replacements which meet the specifications b The MOS FET input transistors QIOI and QI02 are specially selected and matched order only as a plug in...

Страница 22: ...h at VOLTS Multiplier Switch at 1 and the meter zeroed Measurements are with the Model 153 c At times the meter will not zero on any range with the input shorted Adjust the COARSE ZERO Switch Sl02 Fig...

Страница 23: ...nd adjusting the COARSE and MEDIUM ZERO Controls for null If null is not reached check this stage and check for shorts in the base circuit of Ql07 and Ql08 5 Connect Model 153 to the collector of Q109...

Страница 24: ...FIGURE 14 Chassis Side View 18 16...

Страница 25: ...FIGURE 15 Circuit Board pelS3 19 17...

Страница 26: ...FIGURE 16 Printed Circuit Board PC 186 20 18...

Страница 27: ...rr rr 6 0 N a IT IT ct 610C SWITCH 5101 o N rf rf rf r a a FIGURE 17 Switch MEDIUM ZERO S101 ct a ct 19 N IT I D 6 N N N Ii r1 r1 610C SWITCH 5102 FIGURE 18 Switch COARSE ZERO 5102 DECK 2 DECK DECK 2...

Страница 28: ...IO 4 jl We h R IOS I 0 jl aiC h I R IOG jL h dCI07 I RI07 dCI06 6iC h I RIOS UiC H RIO 9 f i C h RIIO J I L f 1 Rill I b 0 NC h R 1I2 i J c P P ocp IJ I C 10Z I I CI03 I j db I IJ CI04 6 db IJ C lOS I...

Страница 29: ...0 CJ CJ 0 Tl m rn ITI n n A A A N VJ 0 0 RI75 I I l RI7 Jl c u R17 3 i R t72 c jl Jl 0 J RI70 RlioS CI23 C 2 o R IGG C 1 2 1 FIGURE 20 Switch Multiplier Sl07 23 21...

Страница 30: ...CALIBRATION Zero whenever other adjustments are made a To set the mechanical Zero meter turn the METER Switch to METER OFF and set the mechanical Zero Meter adjustment for zero meter reading top scale...

Страница 31: ...rent of the Model 6l0C set the front panel controls to ZERO CRECK Button Multiplier Switch Range Switch FEEDBACK Switch METER Switch LOCK 003 10 11 AMPERES FAST Cap the INPUT Receptacle and unlock the...

Страница 32: ...from 10 6 to 10 8 am ere ranges and 4 of full scale from the 10 9 to 10 1 ampere ranges For the ranges above 10 5 ampere construct a current source whose output is 0 2S Set the Model 610C FEEDBACK Swi...

Страница 33: ...t range 1 Set the Signal Generator to 20 Hz at minimum output Release the Electrometer ZERO CHECK Button and increase the Generator output to obtain 2 volt rms at the Model GlOC output 2 Set the Gener...

Страница 34: ...26...

Страница 35: ...ETB F Fig GCb k M or meg Mfg Circuit DesiS C10l C102 C103 C104 C10S C106 C107 C108 C109 CllO Clll Cll2 Cll3 Abbreviations and Symbols ampere Ceramic disc Composition Deposited Carbon Electrolytic Alum...

Страница 36: ...Pilot Light Mfg No 53 Pilot Light Holder Mfg No 203CE Fuse slow blow 1 4A 3AG Mfg No 313 250 Fuse slow blow 1 8A 3AG Mfg Type MDL Fuse holder Mfg No 342012 Receptacle UHF INPUT Mfg No 6804 F P1ug UHF...

Страница 37: ...t Assembly Zero Check Button 80164 80164 KN 47 16988A 1 3 14 S104 Slide Switch 117 234V 80164 SW 151 2 4 S105 Slide Switch FEEDBACK 80164 SW 45 1 3 14 S106 Rotary Switch less components RANGE Rotary S...

Страница 38: ...Cb 07716 DCC RI2 68K 17 R126 249 n 1 1 2 W DCb 91637 DCF 1 2 249 Rl2 249 18 R127 249 n 1 1 2 W DCb 91637 DCF 1 2 249 R12 249 18 R128 249 n 17o 1 2 W DCb 91637 DCF 1 2 249 R12 249 18 R129 249 n 1 1 2 W...

Страница 39: ...WI 01686 7009 R95 9 1 20 R174 2 73 n 1 1 4 I WI 01686 7009 R95 2 73 20 R175 91 Q 1 1 4 I WI 01686 7009 R95 91 20 R176 8 6 kn 1 1 2 I DCb 07716 ncc R12 S 6K 14 RI77 20 kn 10 1 5W WllVar 71450 AI RP34 2...

Страница 40: ...l I 17076B Handle Assembly I HH 18 2 Screw t 1 8 32xl 4 slotted 4 D8 32xl 4 slotted P li 3 Bottom Cover Assembly I 21540C Bottom Cover Sheet Metal I 2154lC Tilt Bail I 15004B Support Assembly Right I...

Страница 41: ..._ v 3 FIGURE 21 Chassis Assembly 0 1r 11 35 33...

Страница 42: ...N Wesson Miss 39191 44655 Ohmite Mfg Co 3601 Howard St Skokie Ill 60476 56289 Sprague Electric Co North Adams Mass 58474 Superior Electric Co The 383 Middle St Bristol Conn 06012 63060 Victoreer 0 tr...

Страница 43: ......

Страница 44: ...Landsberger Strasse 65 D 82110 Germering 089 84 93 07 40 Fax 089 84 93 07 34 GREAT BRITAIN Keithley Instruments Ltd The Minster 58 Portman Road Reading Berkshire RG30 1EA 0118 9 57 56 66 Fax 0118 9 5...

Отзывы: