In this section: 

Cables ...................................................................................... 4-1 

External factors ........................................................................ 4-1 

High impedance measurements ............................................... 4-2 

Cables 

Shielded cables should be used with the 3761. The shield should be connected to circuit guard, 

analog LO. This helps prevent excessive radiation from the cables from interfering with any 

equipment. The cable rating must not exceed the maximum allowable signal levels as defined in the 

3761 card specifications. For more information on cable requirements, go to the 

Keithley Instruments 

website

 (

http://www.tek.com/keithley

) and search for 

Switching Handbook

 or 

Low Current 

Measurements Application Note.

 

The test system response is affected by cable capacitance. This should be considered when the 

sources are connected to the 3761. Use a feedback ammeter for currents below 10

-5 

A to increase 

measurement speed and decrease the effects of cable capacitance measurement. 

External factors 

The 3761 is specified for a resistive load. However, external circuit capacitances and inductances can 

cause excessive currents and voltages across the relay contacts. The currents and voltages must be 

limited to within the maximum allowable signal levels. 

Capacitance:

  Charged capacitances in the circuit can cause excessive currents (surge currents)

to flow through the relay contacts if the source is shorted out or if the output is switched to

capacitive loads. This excessive current can weld the relay contacts together. Therefore, use

series resistors, where needed, to limit the maximum current to within the contact ratings. The

maximum signal levels are stated in the specifications.

Inductance:

  The inductance of cables can produce high voltage and arcing across relay

contacts when switching. The peak transients must be limited to within the maximum signal levels

as defined in the 3761 specifications.

When a channel is opened or closed, there is a charge transfer in the picocoulomb range. This is due 

to the mechanical release or closure of the contacts, the contact-to-coil capacitance, and the stray 

capacitance between the signal and relay drive lines. The charge transfer causes a current pulse, and 

the effect on the signal depends on the magnitude of the source being measured. 
Extreme environmental conditions can cause the offset current to exceed the 1 pA specification. If the 

card has been exposed to high humidity or temperature (for example, during shipping), stabilize the 

board within the specified environmental limits for 24 hours. 

Section 4 

Operation considerations 

Содержание 3761

Страница 1: ...User s Manual 3761 900 01 Rev B January 2018 P3761 900 01B 3761 900 01B www tek com keithley Model 3761 10 Channel Low Current Multiplexer Card...

Страница 2: ...s is strictly prohibited These are the original instructions in English All Keithley Instruments product names are trademarks or registered trademarks of Keithley Instruments Other brand names are tra...

Страница 3: ...d to mains These instruments will be marked as category II or higher Unless explicitly allowed in the specifications operating manual and instrument labels do not connect any instrument to mains Exerc...

Страница 4: ...lains hazards that could damage the instrument Such damage may invalidate the warranty The CAUTION heading with the symbol in the user documentation explains hazards that could result in moderate or m...

Страница 5: ...nstallation 2 2 Web browser support 2 3 General operation 3 1 General operation 3 1 Safety hazards 3 1 Current and voltage switching applications 3 2 Operation considerations 4 1 Cables 4 1 External f...

Страница 6: ...re 1 Model 3761 10 channel Low Current Multiplexer Card The 3761 has two output connectors which allow you to add additional cards Each additional card increases the capacity of the multiplexer by ten...

Страница 7: ...nded warranty Additional years of warranty coverage are available on many products These valuable contracts protect you from unbudgeted service expenses and provide additional years of protection at a...

Страница 8: ...http www tek com keithley General information General ratings for the 3761 are listed in the following table Category Specification Voltage rating 30 V Current rating 100 mA Input and output connecti...

Страница 9: ...lified personnel Do not perform these procedures unless you are qualified Failure to recognize and observe standard safety precautions could result in personal injury or death due to electric shock Un...

Страница 10: ...puts are located on the front of the card Figure 2 Model 3761 card connectors While SMA connectors are not typically used in low current applications the connectors on the 3761 are compact Teflon insu...

Страница 11: ...last inch or so press in firmly to mate the card connector to the mainframe connector 5 On each side of the card there is a mounting screw Tighten these two screws to secure the card to the mainframe...

Страница 12: ...and any unprotected leads wiring with double insulation for 500 volts Category I Follow these recommendations before and after applying power to the equipment 1 Do not exceed the 3761 maximum allowab...

Страница 13: ...cations For current switching applications each channel on the 3761 has a pair of single pole single throw SPST normally open NO relays The relay pair switches signal HI between signal LO analog groun...

Страница 14: ...a low impedance picoammeter When a channel is not selected for measurement the current path is maintained through that channel s analog grounding relay which is automatically updated with the channel...

Страница 15: ...voltsampsmode to any value other than 0 or 1 will result in the following error 222 Parameter data out of range slot slot_3761 voltsampsmode 2 setting voltsampsmode on a card that does not support it...

Страница 16: ...apacitance Charged capacitances in the circuit can cause excessive currents surge currents to flow through the relay contacts if the source is shorted out or if the output is switched to capacitive lo...

Страница 17: ...nt Multiplexer Card User s Manual 4 2 3761 900 01 Rev B January 2018 High impedance measurements Because of the high impedance circuits on the 3761 this card contains no user serviceable components an...

Страница 18: ...f the scanner card include Low offset current error 1 pA specified 30 fA typical Switching of AC or DC signals up to 100 mA SMA input and output connectors Two outputs for connecting multiple 3761 car...

Страница 19: ...h up to 1 MHz Maximum switched and carry current and voltage levels are 100 mA and 30 V peak using resistive loads Direct measurement of capacitor leakage or insulation resistance If you want to measu...

Страница 20: ...citor under test 3 When the capacitor is charged close the applicable channel of the 3761 to start the measurement The current is measured at IM as shown in the previous figure To maintain the voltage...

Страница 21: ...esistance and an individual discharge resistor The discharge resistors are used to drain the electric charge from the capacitors when the test completes Using the indirect method the Model 3762 can be...

Страница 22: ...rd from the outputs when all channels are open and closes when any channel closes This isolation relay configures itself automatically when assembling a test system using multiple 3761 cards The 3706A...

Страница 23: ...Section 5 Typical applications Model 3761 10 Channel Low Current Multiplexer Card User s Manual 5 6 3761 900 01 Rev B January 2018 Figure 8 20 channel scanner system with two 3761 cards in ampere mode...

Страница 24: ...xternal factors 4 1 G General information 1 3 General operation 3 1 H Handling precautions 2 1 High impedance measurements 4 2 I Indirect measurement of capacitor leakage or insulation resistance 5 2...

Страница 25: ...names are the property of Keithley Instruments All other trademarks and trade names are the property of their respective companies Keithley Instruments Corporate Headquarters 28775 Aurora Road Clevel...

Отзывы: