Model 3761 10-Channel Low Current Multiplexer Card User's Manual 

Section 2: Installation and connections 

3761-900-01 Rev. B

 

/

 

January 2018 

2-3

Perform the following steps to install a switching card into the instrument mainframe: 
1. Turn the instrument off and disconnect the power line cord and any other cables connected to the

rear panel.

2. Position the instrument so that you are facing the rear panel.
3. Remove the slot cover plate from the desired mainframe slot. Retain the plate and screws for

future use.

4. With the top cover of the switching card facing up, align the card's edge into the slot's card guide

and slide in the card. For the last ¼ inch or so, press in firmly to mate the card connector to the

mainframe connector.

5. On each side of the card, there is a mounting screw. Tighten these two screws to secure the card

to the mainframe. Do not overtighten.

6. Reconnect the power line cable and any other cables to the rear panel, then turn the instrument

on.

7. Press the

 SLOT

 key to see the model numbers, description, and the firmware revision of the

installed switching cards, along with the mainframe firmware and DMM (if present).

Item shipped may vary from model pictured here. 

Figure 3: 3761 card installation 

Item

 

Description

 

Card guide (part of the mainframe) 

Card edge (part of the card) 

Card 

Mounting screws (part of the card) 

Web browser support 

Web browser support for the 3761 10-Channel Low Current Multiplexer Card is not supported with the 

3706A Switch System/DMM. 

Содержание 3761

Страница 1: ...User s Manual 3761 900 01 Rev B January 2018 P3761 900 01B 3761 900 01B www tek com keithley Model 3761 10 Channel Low Current Multiplexer Card...

Страница 2: ...s is strictly prohibited These are the original instructions in English All Keithley Instruments product names are trademarks or registered trademarks of Keithley Instruments Other brand names are tra...

Страница 3: ...d to mains These instruments will be marked as category II or higher Unless explicitly allowed in the specifications operating manual and instrument labels do not connect any instrument to mains Exerc...

Страница 4: ...lains hazards that could damage the instrument Such damage may invalidate the warranty The CAUTION heading with the symbol in the user documentation explains hazards that could result in moderate or m...

Страница 5: ...nstallation 2 2 Web browser support 2 3 General operation 3 1 General operation 3 1 Safety hazards 3 1 Current and voltage switching applications 3 2 Operation considerations 4 1 Cables 4 1 External f...

Страница 6: ...re 1 Model 3761 10 channel Low Current Multiplexer Card The 3761 has two output connectors which allow you to add additional cards Each additional card increases the capacity of the multiplexer by ten...

Страница 7: ...nded warranty Additional years of warranty coverage are available on many products These valuable contracts protect you from unbudgeted service expenses and provide additional years of protection at a...

Страница 8: ...http www tek com keithley General information General ratings for the 3761 are listed in the following table Category Specification Voltage rating 30 V Current rating 100 mA Input and output connecti...

Страница 9: ...lified personnel Do not perform these procedures unless you are qualified Failure to recognize and observe standard safety precautions could result in personal injury or death due to electric shock Un...

Страница 10: ...puts are located on the front of the card Figure 2 Model 3761 card connectors While SMA connectors are not typically used in low current applications the connectors on the 3761 are compact Teflon insu...

Страница 11: ...last inch or so press in firmly to mate the card connector to the mainframe connector 5 On each side of the card there is a mounting screw Tighten these two screws to secure the card to the mainframe...

Страница 12: ...and any unprotected leads wiring with double insulation for 500 volts Category I Follow these recommendations before and after applying power to the equipment 1 Do not exceed the 3761 maximum allowab...

Страница 13: ...cations For current switching applications each channel on the 3761 has a pair of single pole single throw SPST normally open NO relays The relay pair switches signal HI between signal LO analog groun...

Страница 14: ...a low impedance picoammeter When a channel is not selected for measurement the current path is maintained through that channel s analog grounding relay which is automatically updated with the channel...

Страница 15: ...voltsampsmode to any value other than 0 or 1 will result in the following error 222 Parameter data out of range slot slot_3761 voltsampsmode 2 setting voltsampsmode on a card that does not support it...

Страница 16: ...apacitance Charged capacitances in the circuit can cause excessive currents surge currents to flow through the relay contacts if the source is shorted out or if the output is switched to capacitive lo...

Страница 17: ...nt Multiplexer Card User s Manual 4 2 3761 900 01 Rev B January 2018 High impedance measurements Because of the high impedance circuits on the 3761 this card contains no user serviceable components an...

Страница 18: ...f the scanner card include Low offset current error 1 pA specified 30 fA typical Switching of AC or DC signals up to 100 mA SMA input and output connectors Two outputs for connecting multiple 3761 car...

Страница 19: ...h up to 1 MHz Maximum switched and carry current and voltage levels are 100 mA and 30 V peak using resistive loads Direct measurement of capacitor leakage or insulation resistance If you want to measu...

Страница 20: ...citor under test 3 When the capacitor is charged close the applicable channel of the 3761 to start the measurement The current is measured at IM as shown in the previous figure To maintain the voltage...

Страница 21: ...esistance and an individual discharge resistor The discharge resistors are used to drain the electric charge from the capacitors when the test completes Using the indirect method the Model 3762 can be...

Страница 22: ...rd from the outputs when all channels are open and closes when any channel closes This isolation relay configures itself automatically when assembling a test system using multiple 3761 cards The 3706A...

Страница 23: ...Section 5 Typical applications Model 3761 10 Channel Low Current Multiplexer Card User s Manual 5 6 3761 900 01 Rev B January 2018 Figure 8 20 channel scanner system with two 3761 cards in ampere mode...

Страница 24: ...xternal factors 4 1 G General information 1 3 General operation 3 1 H Handling precautions 2 1 High impedance measurements 4 2 I Indirect measurement of capacitor leakage or insulation resistance 5 2...

Страница 25: ...names are the property of Keithley Instruments All other trademarks and trade names are the property of their respective companies Keithley Instruments Corporate Headquarters 28775 Aurora Road Clevel...

Отзывы: