Model 2750 Multimeter/Switch System User’s Manual
Limits and Digital I/O
9-11
Setting digital output
The OUTPUT menu (shown in
) is used to control and configure digital outputs.
Menu items for the digital output include:
•
DOUTPUT
— Use to enable (ON) or disable (OFF) the digital outputs.
•
PULSE
— Use to enable (YES) or disable (NO) the pulse option for the digital
outputs.
NOTE
The factory default pulse time is 2ms maximum. Using remote programming,
pulse time can be set from 0.001 to 99999.999 sec. It cannot be set from the front
panel.
•
LSENSE
— Use to select the logic sense; active HIGH or active LOW. With active
high selected, an output will be at approxi5V when a reading is at or
exceeds the limit. Conversely, with active low selected, an output will be at 0V
when a reading reaches or exceeds the limit.
•
MASTR LATCH
— Use to enable (Y) or disable (N) the master limit latch. When
enabled, the master limit remains latched when a reading limit is reached or
exceeded. When disabled, the master limit line releases immediately when the
reading is inside all four limits.
Table 9-1
OUTPUT menu
Menu item
Setting
Description
DOUTPUT
ON or OFF
Enable/disable digital outputs.
PULSE
YES or NO
Enable/disable digital pulse
output.
LSENSE
HIGH or LOW
Select logic sense.
BEEP
NEVER, INSIDE, or OUTSIDE Set beeper for limits
, for
details).
MASTR LATCH
Y or N
Enable/disable master limit latch.
2750-900-01.book Page 11 Wednesday, August 3, 2011 7:56 AM
Содержание 2750
Страница 3: ...2750 900 01 book Page 2 Wednesday August 3 2011 7 56 AM...
Страница 7: ...2750 900 01 book Page 4 Wednesday August 3 2011 7 56 AM...
Страница 17: ...2750 900 01 book Page 10 Wednesday August 3 2011 7 56 AM...
Страница 204: ...Model 2750 Multimeter Switch System User s Manual Buffer 6 15 2750 900 01 book Page 15 Wednesday August 3 2011 7 56 AM...
Страница 205: ...6 16 Buffer Model 2750 Multimeter Switch System User s Manual 2750 900 01 book Page 16 Wednesday August 3 2011 7 56 AM...
Страница 324: ...12 CommonCommands 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Страница 332: ...13 SCPISignalOriented MeasurementCommands 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Страница 348: ...15 SCPIReferenceTables 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Страница 374: ...B Model7700ConnectionGuide 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Страница 386: ...C StatusandErrorMessages 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Страница 394: ...D SignalProcessing SequenceandDataFlow 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Страница 408: ...E MeasurementConsiderations 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Страница 428: ...G IEEE 488BusOverview 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...