Model 2701 Instrument Networking Instruction Manual
Network Connection
2-3
Direct connection
This section contains information on connecting the Model 2701 directly to the Ethernet
port on a PC (see
) using an Ethernet cross over cable. This setup provides a one-to-
one connection between the PC and the Model 2701. When using this connection scheme, the
TCP/IP settings for the PCs Ethernet card and the Model 2701 will need to be set manually (see
).
NOTE
A different IP addresses is required for the Model 2701 and the network interface
card.
Figure 2-1
PC to Single Model 2701
Procedure
NOTE
A direct connection requires use of a cross over cable (not the straight-thru cable
that connects a PC’s Ethernet card to a network hub).
1.
Connect one end CAT 5 cross over cable to Network Interface Card in PC.
2.
Connect other end of CAT 5 cross over cable to Network Interface port on the
Model 2701.
3.
Table 2-2
Additional component requirements for direct connection
Quantity
Component
Note
1
Network Interface Card
Properly installed in PC.
1
CAT 5 Cross over cable
3 meter cross over cable included.
Cross over
cable
Up to 100m