10
Front-panel operation
READ?
‘ Trigger 5 current measurement conversions
and return the average of those 5
conversions. The average of the 5 readings
is displayed on the front panel.
Charger
channel (#2)
SOUR2:VOLT 5
‘ Set output voltage to 5 V.
SENS2:CURR:RANG:AUTO ON
‘ Enable auto range for current.
SOUR2:CURR 750e-3
‘ Set current limit to 750 mA.
SOUR2:CURR:TYPE LIM
‘ Select LIM mode for current limit.
SENS2:FUNC ‘VOLT’
‘ Select the voltage measurement function.
SENS2:NPLC 4
‘ Set integration rate to 4 PLC.
SENS2:AVER 4
‘ Set average reading count to 4.
OUTP2 ON
‘ Turn on the power supply output.
READ2?
‘ Trigger 4 voltage measurement conversion and
return the average of those 4 conversions.
SENS2:FUNC ‘CURR’
‘ Select current measurement function.
READ2:ARR?
‘ Trigger 4 current measurement conversions
and return all 4 conversions. The average of
the 4 readings is displayed on the front
panel.