2260B Series Quick Start Guide
20
OVP Range (V)
3-33
8-88
20-275
20-880
OCP Range (A)
5-79.2
2.7-29.7
0.9-9.9
0.288-3.168
2260B (1080W)
30-108
80-40
250-13
800-4
OVP Range (V)
3-33
8-88
20-275
20-880
OCP Range (A)
5-118.8
4.05-44.55 1.35-14.85 0.432-4.752
Steps
1.
Press the OVP/OCP key. The
OVP/OCP key lights up.
2.
The OVP setting will be displayed on the top
and the OCP setting (or OFF) will be displayed
on the bottom.
V
A
OVP Setting
OCP Setting
OVP Level
3.
Use the Voltage knob to set the
OVP level.
Voltage
OCP Level
4.
Use the Current knob to set the
OCP level, or to turn OCP off.
Current
5.
Press OVP/OCP again to exit. The
OVP/OCP indicator will turn off.
Power switch trip
Set F-95 (Power switch trip) to 1 (to disable the
power switch trip) or to 0 (to enable the power
switch trip) and save.