RTD temperature
1.
Connect the precision decade resistance box (listed in
INPUT and SENSE jacks using 4-wire connections. (See
connecting scheme.)
2.
Configure the Model 2010 temperature function for °C units and RTD temperature
sensor (
α
=0.00385) as follows:
A.
Press SHIFT then SENSOR, and note the unit displays the temperature units:
UNITS: C.
B.
Press ENTER, and note the unit displays the sensor type: SENS: TCOUPLE.
C.
Using the cursor and range keys, set the display as follows: SENS: 4W-RTD.
D.
Press ENTER, and note the unit displays: TYPE: PT100.
E.
Using the cursor and range keys, set the unit for the following display: TYPE:
PT385.
F.
Press ENTER to complete the temperature configuration process.
3.
Select the temperature function by pressing the TEMP key.
4.
Set the decade resistance box to each of the values shown in
, and verify that
the temperature readings are within the required limits.
Table 1-8
Four-wire RTD temperature verification reading limits
Applied resistance*
Reading limits (°C)
(1 year, 18°C–28°C)
22.80
Ω
100.00
Ω
313.59
Ω
-190.14 to -189.86°C
-0.08 to +0.08°C
599.86 to 600.14°C
* Based on a = 0.00385. See text.
Performance Verification
1-13
Содержание 2010
Страница 14: ...1 Performance Verification 1 Performance Verification...
Страница 28: ...2 Calibration 2 Calibration...
Страница 53: ...3 Routine Maintenance 3 Routine Maintenance...
Страница 56: ...4 Troubleshooting 4 Troubleshoot ing...
Страница 74: ...5 Disassembly 5 Disassembly...
Страница 83: ...CONFIDENTIAL PROPRIETARY...
Страница 84: ......
Страница 85: ......
Страница 86: ......
Страница 87: ...6 Replaceable Parts 6 Replaceable Parts...
Страница 98: ...CONFIDENTIAL PROPRIETARY...
Страница 99: ...CONFIDENTIAL PROPRIETARY...
Страница 100: ......
Страница 101: ...A Specifications A Specifications...
Страница 107: ...B Calibration CommandReference B Calibration CommandRef erence...
Страница 125: ...C Calibration Program C Calibration Program...