General program instructions
1.
With the power off, connect the Model 2010 and the calibrator to the IEEE-488 interface
of the computer. Be sure to use shielded IEEE-488 cables for bus connections.
2.
Turn on the computer, the Model 2010, and the calibrator. Allow the Model 2010 and the
calibrator to warm up for at least two hours before performing calibration.
3.
Make sure the Model 2010 is set for a primary address of 16. (Use the front panel GPIB
key to check or change the address.)
4.
Make sure the calibrator primary address is at its factory default setting of 4.
5.
Make sure that the computer bus driver software (CECHP.EXE) is properly initialized.
6.
Enter the QBasic editor, and type in the appropriate program below. Check thoroughly
for errors, then save it using a convenient filename.
NOTE
The program assumes a default calibration code of KI002010. If the calibration code
has been changed, modify the :CAL:PROT:CODE parameter accordingly.
7.
Run the program, and follow the prompts on the screen to perform calibration.
Calibration Program
C-3
Содержание 2010
Страница 14: ...1 Performance Verification 1 Performance Verification...
Страница 28: ...2 Calibration 2 Calibration...
Страница 53: ...3 Routine Maintenance 3 Routine Maintenance...
Страница 56: ...4 Troubleshooting 4 Troubleshoot ing...
Страница 74: ...5 Disassembly 5 Disassembly...
Страница 83: ...CONFIDENTIAL PROPRIETARY...
Страница 84: ......
Страница 85: ......
Страница 86: ......
Страница 87: ...6 Replaceable Parts 6 Replaceable Parts...
Страница 98: ...CONFIDENTIAL PROPRIETARY...
Страница 99: ...CONFIDENTIAL PROPRIETARY...
Страница 100: ......
Страница 101: ...A Specifications A Specifications...
Страница 107: ...B Calibration CommandReference B Calibration CommandRef erence...
Страница 125: ...C Calibration Program C Calibration Program...