SCPI subsystem commands
Detailed information an all SCPI commands are covered in the User’s Manual. In this section,
only the command used for basic operation are summarized.
Defaults
:SYSTem
SYSTem Subsystem
:PREset
Set the 2002 to the system present defaults.
Some of the selected defaults include: DCV, continu-
ous triggers, normal speed (1.0 PLC), auto range and
auto filter.
:POSetup <name>
Select power-on setups:
RST
,
PRESet
or
SAV0
through
SAV9
. Instrument will power up to the
*RST
or system
present defaults, or one of the
*SAV0
–
*SAV9
setups.
Fresh readings
[:SENSe[1]]
SENSe Subsystem
:DATA:FRESh?
Requests a “fresh” reading. This command will not re-
quest the same reading twice. If a new reading has al-
ready been triggered, this command will wait until the
new reading is available.
:CALCulate[1]
CALCulate Subsystem
:DATA:FRESh?
Requests a “fresh” calculated reading. When CALC1
is enabled, sens:data:fres? is the reading before calcu-
lation and calc:data:fres? is the reading after calcula-
tion. When CALC1 is disabled, the readings are the
same.
Function configuration
[:SENSe[1]]
SENSe Subsystem
:FUNCtion <name>
Select measurement function:
<name>
=
'VOLTage:DC'
DCV function
'VOLTage:AC'
ACV function
'CURRent:DC'
DCI function
'CURRent:AC'
ACI function
'RESistance'
Ω
2 function
'FRESistance'
Ω
4 function
'FREQuency'
FREQ function
'TEMPerature'
TEMP function
3-12
IEEE-488 Bus Operation