Bench defaults
The Model 2002 can save one, v e, or ten user setups in non-volatile memory, depending on
the installed memory option. You can select one of the user setups as the power-on default, or
have the instrument power up to either of the two factory defaults (optimized for “BENCH” or
“GPIB” operation).
Since the basic measurement procedures in this manual assume the BENCH defaults, reset
the instrument from the SAVESETUP item of the MAIN MENU by performing the following
steps:
1.
Press the MENU key to display the main menu:
MAIN MENU
SAVESETUP GPIB CALIBRATION
TEST LIMITS STATUS-MSG GENERAL
2.
If the SAVESETUP option is not blinking, press the
key until it is blinking and then
press ENTER to view the setup menu:
SETUP MENU
SAVE RESTORE POWERON RESET
3.
Select the RESET option and press ENTER to view the reset menu:
RESET ORIGINAL DFLTS
BENCH GPIB
4.
Select the BENCH option and press ENTER. The following message is displayed:
RESETTING INSTRUMENT
Enter to confirm; EXIT to abort
5.
Press ENTER to con rm. The display will show DC voltage readings with auto-range
enabled.
Other BENCH default settings for DCV include:
•
Triggers — Continuous measurements.
•
Measurement speed (integration time) — Normal, 1 power line cycle.
•
Digital lter — Advanced, 10 readings, 1% noise tolerance, moving average, enabled.
•
Display resolution — 7.5 digits.
Note: See MENUS in the MEASUREMENT OPTIONS section of this manual for details on
navigating the Main Menu.
Front Panel Operation
1-9