IEEE-488 Bus Command Summary
C-2
:CALibration
:UNPRotected
:PREamp
:G1K <value>
:G1K?
:G10K <value>
:G10K?
:G100K <value>
:G100K?
:Z1K
:Z10K
:Z100K
:FS1K <value>
:FS10K <value>
:FS100K <value>
:DATE Ò<string>Ó
:DATE?
:NDUE Ò<string>Ó
:NDUE?
:CALCulate
:SAVE
:ABORt
Calibration commands.
Commands not protected by CAL switch.
Preamp commands.
Directly enter 2mV gain.
Request 2mV gain constant.
Directly enter 200µV gain.
Request 200µV gain constant.
Directly enter 20µV gain.
Request 20µV gain constant.
Measure 2mV range zero.
Measure 200µV range zero.
Measure 20µV range zero.
Measure 2mV range full-scale <value>= 1E-3 to 2E-3
Measure 200µV range full- scale. <value> = 100E-6 to 200E-6.
Measure 20µV range full- scale. <value > = 10E-6 to 20E-6
Send calibration date.
Request calibration date.
Send calibration due date.
Request calibration due date.
Calculate calibration constants.
Save calibration constants in NVRAM.
Abort calibration procedure.
Notes:
1. Angle brackets (<>) are used to indicate parameter type. Do not include brackets in programming message.
2. Upper-case letters indicate command short form.
3. Brackets ([]) indicate command is optional.
4. Commands must be sent in order listed in Table 5-3.
Table C-1 (Continued)
IEEE-488 bus command summary
Command
Description
Содержание 1801