Installation
2-4
Analog output connections
The analog output terminals, which are located on the
power supply board, provide a 0-2V full scale output for
monitoring devices such as chart recorders.
Since the analog output signal is at a relatively high level
and has low source impedance, the type of wiring used for
connections is not critical. Standard stranded wire of the
type used for DMM test leads should be adequate in most
cases. Figure 2-4 shows typical analog output
connections.
Note that the common-mode current (from LO to earth
ground) of any device connected to the analog output ter-
minals will be added to that of the Model 1801/2001. Typ-
ical common-mode current levels for a DMM or chart
recorder are several micro amps or higher.
Power
Connections
Output Connections
To Preamp
Figure 2-3
Power and output connections
2.3
Power supply card preparation
Power and output connections
Before installing the power supply card in the multimeter,
make sure the power cable is connected to the power con-
nector (see Figure 2-3).
Shield jumper
The shield jumper connects the cable shield and preamp
shell either to earth ground or input LO. This jumper
should be installed as follows:
•
If low common-mode current is more important than
common-mode noise rejection, place the jumper in
the earth ground position.
•
If common-mode noise rejection is more important
than low common-mode current, place the jumper in
the LO position.
Содержание 1801
Страница 1: ...Model 1801Nanovolt Preamp Instruction Manual A G R E A T E R M E A S U R E O F C O N F I D E N C E...
Страница 71: ......
Страница 72: ......
Страница 73: ...A Specifications A 1...
Страница 74: ......
Страница 75: ......
Страница 76: ......