Page 36
DOC0935 (26 July 2022 12:02)
www.keit.co.uk
Back to Table of Contents
Figure 27: IRmadillo on an inclined stand
5.7.1. Safety
WARNING
The spectrometer and inclined stand assembly weighs ~24kg. Use
caution when handling to avoid injury.
5.7.2. Installation
Assemble the inclined stand to the desired height; two positions are possible. For technical
drawings, contact Keit at [email protected].
To install the inclined stand to the instrument, carefully roll the spectrometer over such that
the mounting holes are pointed upward, the instrument should be disconnected from power
coms and purge to avoid damage to the cables or twisting.
Position the inclined stand over the mounting points such that the mounting plate lines up with
the instrument. The spectrometer is located with 4 screws through the corners of the stand
into the mounting holes on the spectrometer.
Return the instrument to the correct orientation and connect the purge, the power cable, and
communications cable/s.