T e c a l  1 4 0 F,  T e c a l  4 2 5 F  a n d  T e c a l  6 5 0 F  O P E R A T O R ’ S  M A N U A L

TECHNICAL INFORMATION

Brief fault finding notes and lists of replacement parts, accessories and
inserts for the Tecal are given in this section.

NOTE THAT THIS EQUIPMENT SHOULD ONLY BE DISMANTLED BY
PROPERLY TRAINED PERSONNEL. 

REMOVING THE OUTER COVERS

OR BASE EXPOSES POTENTIALLY LETHAL MAINS VOLTAGES.

THERE ARE NO OPERATOR SERVICEABLE PARTS WITHIN THE
EQUIPMENT.

General advice

In the unlikely event that you experience any problems with your Tecal which
cannot be easily remedied, you should contact your supplier and return the
unit if necessary. Please include any details of the fault observed and
remember to return the unit in its original packing. The insert must be
removed from the unit and packed separately within the case. Techne
accept no responsibility for damage to units which are not properly packed
for shipping: if in doubt, contact your supplier.

General fault finding

HELP display

If HELP is displayed there is a problem with the internal sensing device.  The
calibrator should be returned to your supplier for repair.

Over-temperature Cut-out

If the over-temperature cut-out trips it can be reset by switching the unit off,
waiting a few seconds,then switing it on again.  If it repeatedly trips, please
contact your local supplier.

Cleaning your Tecal unit

Before cleaning your unit, disconnect from the power supply and allow to
cool to ambient temperature.
You can clean the case of the Tecal with a cloth dipped in water or ethanol
(methanol can also be used). No part of the case or cover should be
immersed in the solvents.
Do not use acetone or abrasive cleaners.
Before using any cleaning or decontamination method except those
recommended in this manual, the responsible body should check with
Techne that the proposed method will not damage the equipment.

Fuses

If neither the power light nor display (on the front panel) is lit, one of the two
fuses may have blown. Make sure there is no external cause (such as a
faulty plug or lead).
Fuses should only be changed by suitably qualified personnel.  If the fuses
blow persistently, a serious fault is indicated and you should return the
calibrator to your supplier for repair.
Never fit a fuse rated higher than the value indicated on the unit, serious
damage or personal injury may result.

Содержание TECHNE Tecal 140F

Страница 1: ...Issue 2 Date of issue 3rd May 2005 Techne is a trademark Techne 2003 Tecal 140F Tecal 425F and Tecal 650F Calibrators Operator s Manual...

Страница 2: ......

Страница 3: ...ion 9 Specification Tecal 140F 10 Specification Tecal 425F 11 Specification Tecal 650F 12 WHEN YOU SWITCH ON 13 Front Panel controls 13 Setting the Units for Temperature 13 Setting the Operating Tempe...

Страница 4: ...PAGE 2 T e c a l 1 4 0 F T e c a l 4 2 5 F a n d T e c a l 6 5 0 F O P E R A T O R S M A N U A L...

Страница 5: ...PAGE 3 T e c a l 1 4 0 F T e c a l 4 2 5 F a n d T e c a l 6 5 0 F O P E R A T O R S M A N U A L NOTES...

Страница 6: ...if required ensuring that the air inlet vents on the underside are free from obstruction 5 Note that the following symbols may be next to the indicator lamps on the front panel of the units and have t...

Страница 7: ...tte d aspiration en s assurant que les trous d a ration situ s sous l appareil ne sont pas obstru s 5 Les symboles ci dessous situ s c t des t moins lumineux sur la face avant de l appareil ont la sig...

Страница 8: ...ungsanleitung komplett bevor Sie dieses Ger t benutzen Warnung HOHETEMPERATURENSINDGEF HRLICH siek nnen dem Bediener ernsthafte Verletzungen zuf gen und brennbare Materialien k nnen sich leicht entz n...

Страница 9: ...e pueden aparecer junto a las luces indicadoras en el panel frontal del equipo tienen los siguientes significados Indicador de potencia Indicador del calor Indicador de sobretemperatura 6 Los s mbolos...

Страница 10: ...obe sizes these are detailed towards the back of this manual Maintaining a set temperature The required temperature is set on the calibrator and the operation of the probe is checked ukAS certificate...

Страница 11: ...ginal packing in case you ever need to return it for service or repair When returning a unit remember to remove the insert from the temperature controlled block The unit must be transported in the ori...

Страница 12: ...2 C 0 36 F Temperature stability after 10 minutes 0 05 C 0 09 F Display resolution 0 01 C 0 1 F Set point display resolution 0 1 C 0 1 F Heat up rate 5 minutes 20 C to 100 C 68 F to 182 F Cool down ra...

Страница 13: ...ability at 200 after 10 minutes 0 03 C 0 054 F Temperature stability at 425 after 10 minutes 0 05 C 0 09 FDisplay resolution 0 01 C 0 1 F Set point display resolution 0 1 C 0 1 F Heat up 12 minutes 20...

Страница 14: ...ormity 1 C 1 8 F Temperature stability after 10 minutes 0 09 C 0 16 F Display resolution 0 01 C 0 1 F Set point display resolution 0 1 C 0 1 F Heat up 35 minutes 20 C to 600 C 68 F to 1112 F Cool down...

Страница 15: ...llows It will display this for 1 second then the actual temperature of the block will be indicated The Front Panel Controls The front panel controls consist of three buttons for controlling the displa...

Страница 16: ...ock Indicator The next indicator shows when there is power to the block When the set temperature is different to the block temperature this indicator will light If the light is on continuously the blo...

Страница 17: ...en them The display may reach temperature in 2 or 3 minutes while the insert may take 5 minutes or more to reach temperature After Use 1 When you have finished heating samples remember that parts of t...

Страница 18: ...d the PC before either unit is powered up otherwise data integrity cannot be guaranteed Once the cable is fitted it does not matter which unit is powered up first ON OFF switch RS232 9 way PC serial p...

Страница 19: ...e calibrator should be returned to your supplier for repair Over temperature Cut out If the over temperature cut out trips it can be reset by switching the unit off waiting a few seconds then switing...

Страница 20: ...requires an insert with a 10mm diameter or larger hole 6103890 RS232 Cable only Inserts Tecal 140F and Tecal 425F Inserts are made from aluminium or aluminium bronze and must be ordered separately fro...

Страница 21: ...lled to suit your application Aluminium Bronze Inserts Immersion Depth 152 4 mm or 6 NUMBER PROBE DIAMETER METRIC FINSABA 5 off 6mm FINSABB 1 off each 10mm 8mm 6mm 4 5mm 3mm FINSABC 2 off 6mm 2 off 10...

Страница 22: ...PAGE 20 T e c a l 1 4 0 F T e c a l 4 2 5 F a n d T e c a l 6 5 0 F O P E R A T O R S M A N U A L NOTES...

Страница 23: ...g the details below UK Office Keison Products P O Box 2124 Chelmsford Essex CM1 3UP England Tel 44 0 330 088 0560 Fax 44 0 1245 808399 Email sales keison co uk Please note Product designs and specific...

Отзывы: