Specification Tecal 140F
The figures stated in the specification are as measured 0-50 mm from the base of the
well of the insert, using an SPRT, at the time of calibration.
Minimum temperature
*45°C below ambient*
81°F below ambient
Typically -20°C in an
Typically -4°F in an
ambient of 25°C
ambient of 77°F
Maximum temperature
140°C
284°F
Temperature accuracy
±0.3°C
±0.54°F
Temperature uniformity
±0.2°C
0.36°F
Temperature stability after 10 minutes
±0.05°C
±0.09°F
Display resolution
0.01°C
0.1°F
Set point display resolution
0.1°C
0.1°F
Heat up rate
5 minutes
20°C to 100°C
68°F to 182°F
Cool down rate
9 minutes
100°C to 0°C
182°F to 32°F
Fan cooling
Automatic
Automatic
Dimensions
Height/WidthDepth
273mm/207mm/289mm
10.7"/8.2"/11.4"
Large well in the block:
Depth/Diameter
130mm/38.2mm
5.1"/1.5"
Weight
11Kg
Options
Central Well for inserts, see the list of inserts page 18
Full bi-directional RS232 Comms Port 9way D type plus software and cable
Soft carry case
Hard carry case
UKAS calibration certificate
*Although temperature can be set to -40°C this can only be achieved in an ambient of 5°C.