Pulsair Intellipuff Tonometer by KEELER
16.1 ELECTROMAGNETIC EMISSIONS
Guidance and manufacturer’s declaration – electromagnetic emissions
The Keeler Pulsair IntelliPuff Tonometer is intended for use in the electromagnetic environment
specified below. The customer or user should assure that it is used in such an environment.
Emissions test
Compliance Electromagnetic environment – guidance
RF emissions CISPR 11
Group 1
The Keeler Pulsair IntelliPuff Tonometer uses RF energy
only for its internal function. Therefore, its RF emissions
are very low and are not likely to cause any interference
in nearby electronic equipment.
RF emissions CISPR 11
Class A
The Keeler Pulsair IntelliPuff Tonometer is suitable for use
in all establishments including domestic establishments
and those directly connected to the public low-voltage
power supply network that supplies buildings used for
domestic purposes.
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations / flicker
emissions IEC 61000-3-3
Complies
16.2 ELECTROMAGNETIC IMMUNITY
Guidance and manufacturer’s declaration – electromagnetic immunity
The Keeler Pulsair IntelliPuff Tonometer is intended for use in the electromagnetic environment
specified below. The customer or user should assure that it is used in such an environment.
Immunity test IEC 55015 Test
level
Compliance
level
Electromagnetic
environment – guidance
Electrostatic
discharge (ESD).
IEC 61000-4-2
± 8 kV contact
± 15 kV air
± 8 kV contact
± 15 kV air
Floors should be wood, concrete
or ceramic tile.
If floors are covered with synthetic
material, the relative humidity
should be at least 30%.
Electrical fast
transient/burst.
IEC 61000-4-4
± 2 kV for power
supply lines
± 1 kV for input/
output lines
± 2 kV for power
supply lines
± 1 kV for input/
output lines
Mains power quality should be
that of a typical commercial or
hospital environment.
Surge.
IEC 61000-4-5
± 1 kV line(s) to
line(s)
± 2 kV line(s) to
earth
± 1 kV line(s) to
line(s)
± 2 kV line(s) to
earth
Mains power quality should be
that of a typical commercial or
hospital environment.
Содержание Pulsair IntelliPuff
Страница 1: ...Non Contact Tonometer INSTRUCTIONS FOR USE EN EL PL NL HU...
Страница 2: ......
Страница 4: ......
Страница 29: ...25 Pulsair Intellipuff KEELER EL Keeler Keeler EP29 32777 Pulsair IntelliPuff Keeler Keeler...
Страница 30: ...26 Pulsair Intellipuff KEELER EL Pulsair IntelliPuff 1 2 3 IOP IOP 3 3 1 1 2 3 4...
Страница 33: ...29 Pulsair Intellipuff KEELER EL 4 5 6 Keeler 7 8 OD OS OD OS 2 1 3 4 5 6 7 8 14 10 12 13 11 9...
Страница 38: ...34 Pulsair Intellipuff KEELER EL 9 STBY OK OD 14 mmHg LED OD OS 0 1 0 1 1 IOP 25 mmHg IOP 25 mmHg 25 15 OD OS...
Страница 39: ...35 Pulsair Intellipuff KEELER EL 10 10 1 20 IOP xx x xx 11 1 2 3 4 12 1 3 2 3 2 1...