Pulsair Intellipuff Tonometer by KEELER
3.2 CLEANING TONOMETER
BODY
Only manual non-immersion cleaning as described should be used for this tonometer. Do not
autoclave or immerse in cleaning fluids. Always disconnect power supply from source before
cleaning.
1. Wipe the external surface with a clean absorbent, non-shedding cloth dampened with
de-ionised water / detergent solution (2% detergent by volume) or water / isopropyl
alcohol solution (70% IPA by volume). Avoid optical surfaces.
2. Ensure that excess solution does not enter the instrument. Use caution to ensure cloth is
not saturated with solution.
3. Surfaces must be carefully hand-dried using a clean non-shedding cloth.
4. Safely dispose of used cleaning materials.
4. POWER SUPPLY ASSEMBLY
4.1 SET PLUG
Replace the blanking plate with the appropriate mains plug adapter if required, or use IEC
60320 TYPE 7 connector (not supplied).
5. WALL MOUNTING
Your Pulsair IntelliPuff is supplied with a sturdy wall mounting bracket.
1. The bracket has four holes allowing it to be securely fixed to an
appropriate wall or vertical surface.
2. Choose carefully the intended location for your IntelliPuff with
particular consideration to health and safety aspects, for example, the
routing of the power lead, and its position in regard to the user and
the patient.
3. Use the wall mounting bracket as a template and clearly mark the
position of the holes in the wall. Ensure that there are no live utilities
where you are to drill.
4. Drill the appropriate size holes for the screws and rawplugs supplied.
5. Attach the plate securely to the wall.
6. Remove labels covering the key holes.
7. Carefully locate the mounting pegs of the plate to the key holes on the back of your
Pulsair IntelliPuff and lower into its final and secure position.
6. TONOMETRY, PRESSURE VARIATIONS IN THE HUMAN EYE
The Keeler Pulsair IntelliPuff Tonometer measures intra-ocular pressure by automatically
releasing a gentle puff of air onto the cornea. This is known as an event.
Содержание Pulsair IntelliPuff
Страница 1: ...Non Contact Tonometer INSTRUCTIONS FOR USE EN EL PL NL HU...
Страница 2: ......
Страница 4: ......
Страница 29: ...25 Pulsair Intellipuff KEELER EL Keeler Keeler EP29 32777 Pulsair IntelliPuff Keeler Keeler...
Страница 30: ...26 Pulsair Intellipuff KEELER EL Pulsair IntelliPuff 1 2 3 IOP IOP 3 3 1 1 2 3 4...
Страница 33: ...29 Pulsair Intellipuff KEELER EL 4 5 6 Keeler 7 8 OD OS OD OS 2 1 3 4 5 6 7 8 14 10 12 13 11 9...
Страница 38: ...34 Pulsair Intellipuff KEELER EL 9 STBY OK OD 14 mmHg LED OD OS 0 1 0 1 1 IOP 25 mmHg IOP 25 mmHg 25 15 OD OS...
Страница 39: ...35 Pulsair Intellipuff KEELER EL 10 10 1 20 IOP xx x xx 11 1 2 3 4 12 1 3 2 3 2 1...