Refer to Ophthalmic instructions for Ophthalmoscope - Otoscope operating 

procedures.

Dispos-A-Spec

To dispense a speculum, simply grasp the end of the required speculum and 

gently pull vertically. When a dispenser tube is empty, reorder speculum using 

the order form EP59-48483. Remove the lid from the unit and refill the required 

tube.

Warranty & Service

No user serviceable parts - all preventative maintenance and servicing must 

only be performed by an authorised Keeler representative.

Your Keeler wall unit is guaranteed for 3 years and will be replaced or repaired 

free of charge subject to the following:-

1. 

Any fault is due to faulty manufacture

2. 

The power supply has been used in compliance with these instructions

3. 

Proof of purchase accompanies any claim.

Technical Information

Input data: 

100-240V - 50/60Hz

Power Consumption: 

2.5 Amps

Operation: 

Continuous

Classification: 

Class II equipment

 

Type B Protection against electric shock

Safety Class: 

Double insulated

Ambient Temperature: 

+10°C to +35°C

Transportation/Storage: 

Ambient temperature -40°C to +70°C

 

Relative humidity of 10% to 100% including 

condensation

Complies to EN60601-1

This equipment may be affected by electromagnetic interference.

Other electrical equipment in the close vicinity may also be affected by the Gen. 

Med. Unit.

If such effects are suspected, switch off the offending equipment.

Содержание 1964-P-2001

Страница 1: ...PLEASE READ AND FOLLOW THESE INSTRUCTIONS CAREFULLY 1964 P 2001 1964 P 2000 ...

Страница 2: ...ontinued use of a damaged product or part may cause injuries Do not immerse the unit in any liquid Do not use unit if damaged and periodically visually inspect for signs of dam age Keeler cord handles attached to a mains unit provide a 3 6v output to the instru ment head Care must be taken to ensure that only instrument heads fitted with a 3 6V bulb are connected Caution Do not fit into damaged el...

Страница 3: ...ower supply to unit and plug in mains fig 1 CHECK fig 2 WALL UNIT DISPENSER TEMPLATE see attached WALL UNIT DISPENSER TEMPLATE Wall Mounting For the Gen Med Unit Dispenser Units drill two holes Ø6mm x 55mm deep and 110mm apart For the Dispenser Unit drill an additional two holes 249mm below the existing holes using template in fig 2 and attached ...

Страница 4: ...WALL UNIT DISPENSER TEMPLATE Scale 1 1 Do not scale ...

Страница 5: ...fig 3 CAUTION 55mm Ø6mm Ø6mm fig 4 fig 5 ...

Страница 6: ...ed see fig 7 To use the required instrument remove the relevant handle from its cradle as shown in fig 8 fig 6 fig 7 Allen Key fig 8 PULL A yellow light LED will illuminate when a cord handle is removed from its cra dle This will occur whether or not an instrument head is fitted When the instrument is no longer required always ensure that the handle is returned correctly to its cradle and that the...

Страница 7: ...harge subject to the following 1 Any fault is due to faulty manufacture 2 The power supply has been used in compliance with these instructions 3 Proof of purchase accompanies any claim Technical Information Input data 100 240V 50 60Hz Power Consumption 2 5 Amps Operation Continuous Classification Class II equipment Type B Protection against electric shock Safety Class Double insulated Ambient Temp...

Страница 8: ...t non shedding cloth dampened with a water detergent solution 2 by volume or water isopropyl alcohol solution 70 by volume 2 Ensure that excess solution does not enter the instrument Use caution to ensure cloth is not saturated with solution 3 Surfaces should be carefully hand dried using a clean non shedding cloth 4 Safely dispose of used cleaning materials ...

Страница 9: ...HONE 0800 521 251 Keeler Instruments Inc 456 Parkway Broomall PA 19008 USA Toll Free 1 800 523 5620 Tel 610 353 4350 Fax 610 353 7814 EP59 48475 Issue A A HALMA GROUP C O M P A N Y DISTRIBUTED BY The CE mark on this product indicates it has been tested to and conforms with the provisions noted within the 93 42 eec Medical Device Directive ...

Отзывы: