10. Maintenance
In Germany, according to MPBetreibV §7 (maintenance), operators of medical devices
are obliged to ensure the safe condition of the medical device throughout its entire
service life.The inspection according to DIN EN 62353 includes the following minimum
requirement:
-
Visual inspection
-
Insulation resistance measurement*
-
Leakage current measurement*
-
Functional test
-
Overall assessment
To ensure safe operation, a visual and functional inspection with electrical testing
must be carried out at least once a year*.
Test the care bed once a year according to DIN EN 62353 (see protocol for tes
-
ting in chapter „Appendix“).
Copy the protocol (see protocol for testing in chapter „Appendix“) and docu
-
ment the test in the copy.
Have the care bed serviced by qualified personnel only. Contact the manufac
-
turer (see chapter „Manufacturer“).
During maintenance and repair work on the care bed or parts of the care bed,
no persons may be in the care bed.
(* Only for electrically adjustable care beds).
12. Repair
Have the care bed repaired by qualified personnel only. Contact the manufac
-
turer (see chapter „Manufacturer“).
Attention
Only original parts and components approved by the manufacturer may be
used.
Electrical components may only be repaired or replaced by authorised spe-
cialist personnel. Contact the manufacturer (see chapter „Manufacturer“).
Содержание NILS
Страница 1: ...Instruction manual Travel Bed NILS...
Страница 2: ...2 Instruction manual travel bed NILS...
Страница 3: ...Instruction manual travel bed NILS 3 KEEP FOR LATER USE READ CAREFULLY...
Страница 5: ...Instruction manual travel bed NILS 5...
Страница 13: ...Instruction manual travel bed NILS 13 3 Scope of delivery Travel bed NILS 9 8 3 2 7 5 1 6 6 4 4 10 11 12 13...