Chapter 10: Sensor Verification
Kaye ValProbe RT User Manual
99
The Verification Results screen will display the Pass / Fail analysis results of the Verification
for each sensor. You can select the setpoints to view the results for each setpoint.
Additional Verification report options:
•
Check detailed sensor readings and IRTD readings using the Sensor Readings
button.
•
Display the sensor and IRTD readings in graphical format
•
Generate a PDF Verification report for all selected verified sensors
•
Change various criteria, in the textboxes
When completed with the Verification Results select
Exit
.
10.6. Reasons for Verification Failure
Verification may fail for some sensors due to one of the following reasons:
•
Failed criteria of allowed Logger deviation
•
Failed Verification due to unstable IRTD and Loggers
•
Logger failed stability criteria and cannot be evaluated
•
Time of stability of IRTD and Logger may not overlap a sufficient time
•
Logger were removed from the bath/calibrator before the setpoint was reached
10.7. Verification Report
Once the process is completed, a Verification report can be generated
Figure 114: Verification Report