6   Preparation methods according to ISO 17664

Note
The following preparation procedure applies for the SONICflex instrument,

torque wrench and nozzle needle.

6.1   Preparations at the site of use

WARNING

Hazard from non-sterile products.

There is a risk of infection from contaminated medical devices.

▶ Observe suitable personal protective measures.

Remove all residual cement, composite or blood without delay.

 

 

 

Preparation methods according to ISO 17664

38

Содержание SONICflex 2003

Страница 1: ...Instructions for use SONICflex 2003 REF 1 000 4246 SONICflex LUX 2003 L REF 1 000 4243...

Страница 2: ...Distributed by KaVo Dental GmbH Bismarckring 39 D 88400 Biberach Tel 49 7351 56 0 Fax 49 7351 56 1488 Manufacturer Kaltenbach Voigt GmbH Bismarckring 39 D 88400 Biberach www kavo com...

Страница 3: ...ription of safety instructions Warning symbol 9 2 1 2 Description of safety instructions Structure 9 2 1 3 Description of safety instructions Description of danger levels 10 2 2 Purpose Proper use 11...

Страница 4: ...ULTIflex and MULTIflex LUX coupling 27 4 3 Check O rings 27 4 4 Check the pressure 28 5 Operation 30 5 1 Attaching SONICflex 30 5 2 Removing SONICflex 32 5 3 Inserting SONICflex tip 33 5 4 Removing SO...

Страница 5: ...eaning Manual cleaning internal 42 6 2 4 Cleaning Automated internal cleaning 44 6 3 Disinfection 45 6 3 1 Disinfection Manual disinfection external 46 6 3 2 Disinfection Manual disinfection internal...

Страница 6: ...Care with KAVOspray 54 6 5 3 Care products and systems Servicing Care of the KaVo SPRAYrotor 55 6 5 4 Care products and systems Servicing Care with KaVo QUATTROcare 56 6 6 Packaging 57 6 7 Sterilisati...

Страница 7: ...aVo quality product By following the instructions below you will be able to work smoothly economically and safely Copyright by KaVo Dental GmbH Symbols Refer to Safety Warning symbol chapter Important...

Страница 8: ...equirements of the applicable EC directive Can be steam sterilised at 134 C 1 C 4 C 273 F 1 6 F 7 4 F Thermodisinfectable Target group This document is intended for dentists and their assistants The s...

Страница 9: ...air Other claims of any nature whatsoever in particular with respect to compensation are excluded In the event of default gross negligence or intent this shall only apply in the absence of mandatory l...

Страница 10: ...ived from manipulations or changes to the product by the customer or a third party Service warranty claims will only be accepted if the product is submitted along with proof of purchase in the form of...

Страница 11: ...symbol 2 1 2 Description of safety instructions Structure DANGER The introduction describes the type and source of the hazard This section describes the potential consequences of non observance The o...

Страница 12: ...d here together with the three levels of danger will help avert property damage and injury CAUTION CAUTION indicates a hazardous situation that can cause damage to property or mild to moderate injurie...

Страница 13: ...tended for dental treatment Any other type of use or alteration to the product is impermissible and can be hazardous The SONICflex can be used in conjunction with KaVo tips for plaque removal prophyla...

Страница 14: ...le health and safety regulations the applicable accident prevention regulations these Instructions for use According to these regulations the user is required to only use equipment that is operating c...

Страница 15: ...d patient Stop working in case of damage irregular noise during operation exces sive vibration unusual build up of heat or if the tip cannot be firmly held The instrument must be cleaned serviced and...

Страница 16: ...ment tip becomes loose Stop work and seek service support CAUTION Premature wear and malfunctioning from improper storage during long periods of nonuse Reduced product life The medical device should b...

Страница 17: ...t injury To ensure proper function the medical device must be set up according to the reprocessing methods described in the KaVo Instructions for Use and the care products and care systems described t...

Страница 18: ...1 Safety instructions SONICflex tips CAUTION Risk of injury and infection when changing the SONICflex tips This can substantially endanger the user To test use and remove the SONICflex tips use a glov...

Страница 19: ...Mat no 1 001 6958 The tips may break from continuous stress or damage such as being drop ped on the floor or being bent Therefore check the tip to see if it is safe before each use by applying a slig...

Страница 20: ...For an additional safety check mechanically apply approx 10 N 1 kg to the tip without operating the device Safety 18...

Страница 21: ...3 Product description SONICflex 2003 Mat no 1 000 4246 SONICflex LUX 2003 L Mat no 1 000 4243 Product description 19...

Страница 22: ...spective application The internal water cooling spray cooling prevents the treatment field from heating up and keeps the treatment surface clean Note The device should be operated with the described o...

Страница 23: ...MULTIflex couplings Oscillation level 1 120 15 m Oscillation level 2 160 15 m Oscillation level 3 240 15 m CAUTION Observe the SONICflex recommended setting Non compliance with the recommendations ca...

Страница 24: ...ce after it has been stored strongly refrigerated This can cause the medical device to malfunction Prior to start up very cold products must be heated to a temperature of 20 C to 25 C 68 F to 77 F Tem...

Страница 25: ...Air pressure 700 hPa to 1060 hPa 10 psi to 15 psi Protect from moisture Product description 23...

Страница 26: ...ylaxis endodontics periodontology surgery and conservative dentistry The amount of water needs to be set on the dental unit so that the instru ment tip sprays the water with the proper oscillating int...

Страница 27: ...the tooth and guided parallel to the tooth The in strument should be moved parallel to the surface of the tooth and not the edge to prevent forming notches in the tooth substance Professionals rec om...

Страница 28: ...e from soiled and moist cooling air Contaminated and moist cooling air can cause malfunctions and lead to premature bearing wear Make sure that the supply of cooling air is dry clean and unconta minat...

Страница 29: ...rew the MULTIflex LUX MULTIflex LED coupling onto the turbine hose and tighten with the wrench 4 3 Check O rings CAUTION Missing or damaged O rings Malfunctions and premature failure Make sure that al...

Страница 30: ...tomatically inside the SONICflex The air consumption is approximately 20 40 NI min Insert the test manometer Mat no 0 411 8731 between the MULTIflex coupling and the SONICflex Pressure displayed Drive...

Страница 31: ...First use 29...

Страница 32: ...SONICflex is tightly seated on the coupling If the SONICflex unintentionally comes off the coupling during treatment it can endanger the patient and user Before each treatment pull on the SONICflex t...

Страница 33: ...precise coupling can destroy the high pressure bulb of a MULTIflex LUX coupling or reduce its life Make sure that the coupling is precise Place the SONICflex precisely on the MULTIflex MULTIflex LUX c...

Страница 34: ...5 2 Removing SONICflex Hold the MULTIflex MULTIflex LUX coupling tightly and pull the SONICflex forward while twisting slightly Operation 32...

Страница 35: ...wrench This may result in a risk of injury for the user When inserting the tip into the torque spanner make sure that the tip s end always faces into the opening of the torque spanner Note In view of...

Страница 36: ...t clockwise The torque wrenches are for changing the working tips of the SONICflex and protect against injury The torque wrench can be screwed in more quickly by holding it at the rear thin grip area...

Страница 37: ...flex is in the holder the torque wrench should be placed on the tip to protect against injury 5 4 Removing SONICflex tip Insert attach the torque wrench in on the SONICflex and unscrew the tip anticlo...

Страница 38: ...ce levels 1 2 and 3 5 6 Regulating spray Turn the spray ring on the MULITflex LUX MULTIflex LUX coupling in order to regulate the water content The water volume can be regu lated by selecting differen...

Страница 39: ...ase the volume 5 7 Check the amount of water CAUTION Overheating of the tooth due to lack of cooling water Thermal damage to the dental pulp Adjust the water amount for the spray cooling to a minimum...

Страница 40: ...torque wrench and nozzle needle 6 1 Preparations at the site of use WARNING Hazard from non sterile products There is a risk of infection from contaminated medical devices Observe suitable personal pr...

Страница 41: ...he like Reprocess the medical device as soon as possible after treatment Remove the tip from the SONICflex using the torque wrench 6 2 Cleaning CAUTION Malfunctions from cleaning in an ultrasonic unit...

Страница 42: ...d Tap water 30 C 5 C 86 F 10 F Brush e g medium hard toothbrush Brush it off under running tap water using for example a medium hard toothbrush Prepare the tip according to the manufacturer s instruct...

Страница 43: ...er mielclear rinsing agent and extends only to the compatibility of materials with respect to KaVo products For program settings as well as cleansers and disinfectants to be used please refer to the I...

Страница 44: ...place it on the corresponding care adapter Press the spray button three times for 2 seconds each time Remove the medical device from the spray attachment and let the cleaner work for one minute Afterw...

Страница 45: ...ourg Belgium Netherlands United Kingdom Denmark Sweden Fin land and Norway In other countries only automated interior cleaning with thermodisinfec tors in accordance with ISO 15883 1 Note Do not place...

Страница 46: ...er mielclear rinsing agent and extends only to the compatibility of materials with respect to KaVo products For program settings as well as cleansers and disinfectants to be used please refer to the I...

Страница 47: ...ion CAUTION Malfunctioning from using a disinfectant bath or disinfectant containing chlorine Defects in the product Only disinfect in a thermodisinfector or manually Preparation methods according to...

Страница 48: ...on material compatibility The microbiological efficacy must be ensured by the disinfectant manufac turer Microcide AF from Sch lke Mayr liquid or cloths FD 322 from D rr CaviCide from Metrex Tools req...

Страница 49: ...instructions for use for the disinfectant 6 3 2 Disinfection Manual disinfection internal The efficacy of manual internal disinfection must be demonstrated by the manufacturer of the disinfection agen...

Страница 50: ...nts at a pH of max 10 e g Miele G 7781 G 7881 validation was performed with the VARIO TD programme neodisher mediclean cleaning agent neodisher Z neu tralisation agent and neodisher mielclear rinsing...

Страница 51: ...then lubricate immediately with care agents from the KaVo care system 6 4 Drying Manual drying Blow off the outside and inside with compressed air until water drops are no longer visible Machine dryin...

Страница 52: ...r com pressed air quality see the Warning under Start up 6 5 Care products and systems Servicing CAUTION Premature wear and malfunctions from improper servicing and care Reduced product life Perform p...

Страница 53: ...se listed as accessories as they were tested for proper use on our products Note If you are bothered by some oil leaking during the treatment the technical set up allows for switching from servicing w...

Страница 54: ...ems Servicing Torque wrench care CAUTION Malfunctions from cleaning in the ultrasonic unit Defects in the torque wrench Do not place the torque wrench in ultrasonic cleaning devices Preparation method...

Страница 55: ...is pressed into the tor que wrench in the slots or grease pockets of the locking springs Then place a small amount of grease on your fingertip and press it into the torque wrench at the indicated loca...

Страница 56: ...are with KAVOspray KaVo recommends servicing the product once per week Remove tip Cover the product with the Cleanpac bag Place the product on the cannula and press the spray button for one second Pre...

Страница 57: ...recommends servicing the product once per week Remove tip Place the product on the appropriate coupling of the KaVo SPRAYrotor and cover it with a Cleanpac bag Servicing the product See also Instruct...

Страница 58: ...Servicing Care with KaVo QUAT TROcare Cleaning and care unit with expansion pressure for effective cleaning and care KaVo recommends servicing the product once per week Remove the tip Preparation met...

Страница 59: ...trument so that the bag is not stretched The quality and use of the sterilised product packaging must satisfy ap plicable standards and be suitable for the sterilisation procedure Individually weld th...

Страница 60: ...ERIclave B 2200 2200 P CAUTION Premature weary and malfunctions from improper servicing and care Reduced product life Before each sterilisation cycle service the medical device with KaVo care products...

Страница 61: ...ely remove the product from the steam steriliser after the sterilisation cycle Note Remove the tips to be sterilized and dry the medical device The KaVo medical device has a maximum temperature resist...

Страница 62: ...s at 134 C 1 C 4 C 273 F 1 6 F 7 4 F Autoclave using the gravitation method at least 10 minutes at 134 C 1 C 4 C 273 F 1 6 F 7 4 F or alternatively at least 60 minutes at 121 C 1 C 4 C 250 F 1 6 F 7 4...

Страница 63: ...Note Observe the expiration date of the sterilised item Preparation methods according to ISO 17664 61...

Страница 64: ...Mat no Instrument stand 2151 0 411 9501 Cellulose pad 100 units 0 411 9862 Cleanpac 10 units 0 411 9691 Torque wrench 1 000 4887 Coupling 1 006 5966 Open ended wrench 0 411 0892 Silicone grease 1 000...

Страница 65: ...lus Spray 2140 P 1 005 4525 No Tip type Mat no 5 Universal scaler Mat no 0 571 5171 6 Crescent scaler Mat no 0 571 5181 7 Perio scaler Mat no 0 571 5191 8 Perio extra long scaler Mat no 0 571 5371 12...

Страница 66: ...at no 0 571 5621 25 Root planer button small right Mat no 0 571 5631 26 Root planer button small uni versal Mat no 0 571 5641 27 Rootplaner button large perio Mat no 0 571 5651 28 Micro torpedo shape...

Страница 67: ...mesial Mat no 1 002 1984 35 Prep CAD CAM distal Mat no 1 002 1986 42 Cariex D 0 8 D 64 Mat no 1 000 7105 43 Cariex D 1 2 D 64 Mat no 1 000 7167 45 Seal conical D 46 Mat no 1 000 2568 48 Clean brush ho...

Страница 68: ...ial Mat no 0 571 7262 52 Prep ceram distal Mat no 0 571 7272 53 Angle distal Mat no 0 571 7302 54 Angle mesial Mat no 0 571 7292 55 Retro anterior tooth Mat no 0 571 7342 56 Retro finder left Mat no 0...

Страница 69: ...Mat no 1 000 5822 68 Endo conical 112 Mat no 1 000 5823 69 Endo button small 117 Mat no 1 000 5827 70 Endo conical 117 Mat no 1 000 5821 71 Cariex TC 1 0 Mat no 1 000 7362 72 Cariex TC 1 4 Mat no 1 00...

Страница 70: ...ler Mat no 1 006 0741 87 Bone saw blade Mat no 1 006 1275 Bone saw blade refill Mat no 1 006 1405 Implant Set Mat no 1 003 8167 Implant Pin Refill Mat no 1 003 8168 Swing Set Mat no 1 006 2045 Swing 0...

Страница 71: ......

Страница 72: ...1 006 1353 kb 20100916 05 en...

Отзывы: