Number of available O-rings: 5
4.4 Check the pressure
A minimum drive pressure of 2.5 bar (36 psi) is required for operating the
SONICflex. A higher drive pressure will be reduced automatically inside the
SONICflex. The air consumption is approximately 20-40 NI/min. Insert the
test manometer (Mat. no. 0.411.8731) between the MULTIflex coupling and
the SONICflex.
Pressure displayed:
▪ Drive air T.R. 2.5 to 4.2 bar (36 to 61 psi)
▪ Return air R.L. < 0.4 bar (6 psi)
▪ Spray air Sp.L. = max. 2 (29) bar (psi)
Spray air not required, however.
▪ - Water W.: 1.0 to 2.0 bar (15 to 29 psi)
First use
28
Содержание SONICflex 2003
Страница 1: ...Instructions for use SONICflex 2003 REF 1 000 4246 SONICflex LUX 2003 L REF 1 000 4243...
Страница 25: ...Air pressure 700 hPa to 1060 hPa 10 psi to 15 psi Protect from moisture Product description 23...