![KaVo SONICflex 1.008.6383 Скачать руководство пользователя страница 15](http://html1.mh-extra.com/html/kavo/sonicflex-1-008-6383/sonicflex-1-008-6383_instructions-for-use-manual_3997551015.webp)
▶ For program settings as well as cleansers and disinfectants to be used, please
refer to the Instructions for Use of the thermodisinfector (complying with max.
pH value of 10).
6.5
Drying
Manual Drying
▶ Blow off the outside and inside with compressed air until water drops are no
longer visible.
Automatic Drying
The drying procedure is normally part of the cleaning program of the thermodisin‐
fector.
▶ Follow the instructions for use of the thermodisinfector.
6.6
Packaging
Note
The sterilisation bag must be large enough for the tip so that the bag is not
stretched.
The quality and use of the sterilised product packaging must satisfy applicable
standards and be suitable for the sterilisation procedure.
Note
The SONICflex can also be sterilised in the sterilisation tray.
▶ Individually weld the medical device in the sterilised item packaging (such as
KaVo STERIclave bags Mat. no. 0.411.9912)!
▶ Tips with short thread and marked A only fit in green mounts.
Tips with long thread fit in blue and green mounts.
6.7
Sterilisation
Sterilisation in a steam steriliser (autoclave) in accordance with EN
13060/ISO 17665-1
CAUTION
Contact corrosion due to moisture.
Damage to product.
▶ Immediately remove the product from the steam steriliser after the sterilisa‐
tion cycle!
Instructions for use For SONICflex tips prep crown round, prep crown round A - REF 1.008.6383,
1.008.6384, prep crown plain, prep crown plain A - REF 1.008.6385, 1.008.6386
6 Reprocessing methods according to ISO 17664 | 6.5 Drying