4
First use
WARNING
Hazard from nonsterile products.
Infection danger to the care provider and patient.
▶ Before first use and after each use, prepare and sterilise the medical device
if needed.
4.1
Inserting the SONICflex tips
CAUTION
Hazard from a tip which has been incorrectly inserted in the torque wrench.
This may result in a risk of injury for the user.
▶ When inserting the tip into the torque spanner, make sure that the tip's end
always faces into the opening of the torque spanner.
▶ Insert the desired tip with the tip end pointing down into the torque wrench and
screw it into the handpiece by turning it clockwise.
The torque wrench is for changing the working tips of the SONICflex and for pro‐
tection against injury. The torque wrench can be screwed in more quickly by hold‐
ing it at the rear, thin grip area ①. The large diameter ② is used for tightening
and removal.
Note
The tip is properly gripped when the torque wrench slips.
Note
When the SONICflex is in the holder, the torque wrench should be placed on the
tip to protect against injury.
4.2
Removing the SONICflex tip
▶ Place the torque wrench on the SONICflex tip, and unscrew it anticlockwise.
Instructions for use For SONICflex tips prep crown round, prep crown round A - REF 1.008.6383,
1.008.6384, prep crown plain, prep crown plain A - REF 1.008.6385, 1.008.6386
4 First use | 4.1 Inserting the SONICflex tips
8/15