Brush the SONICflex tip under running tap water using, e.g., a toothbrush of me‐
dium hardness. Clean the cone of the tip with a rubber eraser. If necessary, clear
the water passage at the tip with a nozzle needle.
6.3.2
Automated external cleaning
KaVo recommends thermodisinfectors in accordance with EN ISO 15883 that are
operated with alkaline cleaning agents at a pH value of max. 10 (e. g. Miele G
7781 / G 7881 – validation was performed with the "VARIO-TD" programme,
"neodisher
®
mediclean" cleaning agent, "neodisher
®
Z" neutralisation agent and
"neodisher
®
mielclear" rinsing agent and extends only to the compatibility of
materials with respect to KaVo products).
▶ For program settings as well as cleansers and disinfectants to be used, please
refer to the Instructions for Use of the thermodisinfector (complying with max.
pH value of 10).
6.3.3
Manual internal cleaning
To effectively set up, the inside of the machine must be cleaned automatically in a
cleaning and disinfection unit in accordance with ISO 15883-1.
(The interior of this product is not to be cleaned manually).
6.3.4
Automated internal cleaning
Note
Before setting up, screw the SONICflex tips clockwise onto the SONICflex instru‐
ment.
KaVo recommends thermodisinfectors in accordance with EN ISO 15883 that are
operated with alkaline cleaning agents at a pH value of max. 10 (e. g. Miele G
7781 / G 7881 – validation was performed with the "VARIO-TD" programme,
"neodisher
®
mediclean" cleaning agent, "neodisher
®
Z" neutralisation agent and
"neodisher
®
mielclear" rinsing agent and extends only to the compatibility of
materials with respect to KaVo products).
▶ For program settings as well as cleansers and disinfectants to be used, please
refer to the Instructions for Use of the thermodisinfector (complying with max.
pH value of 10).
6.4
Disinfection
CAUTION
Malfunctioning from using a disinfectant bath or disinfectant containing chlorine.
Defects in the product.
▶ Only disinfect in a thermodisinfector or manually.
Instructions for use For SONICflex tips prep crown round, prep crown round A - REF 1.008.6383,
1.008.6384, prep crown plain, prep crown plain A - REF 1.008.6385, 1.008.6386
6 Reprocessing methods according to ISO 17664 | 6.4 Disinfection